Abstract
Objectives:
in vitro assessment of the reproducibility and the optimal separation and position of the optodes in continuous wave (CW-) NIRS measurement of local inhomogeneities in absorption and/or scattering.
Methods:
a CW- NIRS system (OXYMON) was used with laser diodes at wavelengths of 767 nm, 845 nm, 905 nm, 945 nm and 975 nm. For practical considerations (dimensions of neonatal head) the measurements were performed on a cylindrical tissue-equivalent phantom (70 mm diameter of base material with
Results:
reproducibility of repeated measurements (
Conclusions:
when considering the phantom as a realistic geometrical model for the neonatal head, it can be concluded that the optode position at 90° angle would be optimal for detecting an inhomogeneity at 15 mm depth, i.e. the location of the periventricular white matter. Since the rods are relatively strongly different from the base material the question remains to be answered whether local ischemia, which might lead to irreversible brain damage, can be detected by CW-NIRS