Abstract
Background:
The wearable cardioverter-defibrillator is a treatment option for patients at temporarily high risk of sudden cardiac death or in whom implantation of a cardioverter-defibrillator is temporarily not possible.
Objectives:
The aim of this study was to provide real-world data on patients receiving this therapy in a nurse-based wearable cardioverter-defibrillator training programme.
Methods:
A registry including all patients prescribed with a wearable cardioverter-defibrillator in Austria between 2010 and 2016. Overall, 448 patients received a wearable cardioverter-defibrillator in 48 centres. Patients received structured nurse-based wearable cardioverter-defibrillator educational initial training followed by remote monitoring.
Results:
Main indications were: severe non-ischaemic cardiomyopathy (21%); recent myocardial infarction and percutaneous coronary intervention (20%); and stable coronary artery disease with percutaneous coronary intervention/coronary artery bypass grafting (14%). Eleven patients (2.5%) received 22 appropriate wearable cardioverter-defibrillator shocks. Two patients (0.4%) received three inappropriate shocks. The risk of sudden cardiac death varied between different aetiologies. Eight out of 11 (73%) patients received their first wearable cardioverter-defibrillator shock within 30 days. The main reasons for termination of the wearable cardioverter-defibrillator therapy were implantable cardioverter-defibrillator implantation (55.5%) and improvement of left ventricular ejection fraction to more than 35% (33%).
Conclusion:
The wearable cardioverter-defibrillator is an effective and safe treatment option in patients at either transiently elevated risk of ventricular tachycardia/ventricular fibrillation or mandated postponed implantable cardioverter-defibrillator implantation, with a 2.5% shock rate over a median 54 days wearable cardioverter-defibrillator treatment period. However, both the wearable cardioverter-defibrillator shock rate and implantable cardioverter-defibrillator implantation rate vary widely depending on the wearable cardioverter-defibrillator indication. Nurse-based wearable cardioverter-defibrillator training is associated with high patient adherence, with a median wearing duration per day of 23.5 (1–24) hours.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.