During commissioning of the medical gas system of a new hospital, fourteen connection defects were found; six of these were potentially lethal cross-connections. A major contaminant was also detected throughout the medical gas system.
The methods of testing a new medical gas system described in this report highlight the need for up-grading of existing standards.
References
1.
Australian Standard CB4 (1969): SAA gas cylinders code. Published by the Standards Association of Australia, Standards House, 80 Arthur Street, North Sydney, N.S.W., 2060, Australia.
2.
Australian Standard SAA 1345 (1972): The identification of piping, conduits and ducts. Published by the Standards Association of Australia, Standards House, 80 Arthur Street, North Sydney, N.S.W., 2060, Australia.
3.
Australian Standard 1169 (1973): SAA medical agents and gases safety code. Published by the Standards Association of Australia, Standards House, 80 Arthur Street, North Sydney, N.S.W., 2060, Australia.
4.
British Standard 1319 (1955): Medical gas cylinders and anaesthetic apparatus. Published by British Standards Institute, British Standards House, 2 Park Street, London, W1.
5.
Canadian Standard Z305.1 (1975): Non-flammable medical gas piping systems.Canadian Standards Association, 178 Rexdale Boulevard, Rexdale, Ontario, Canada, M9W 1R3.
6.
Clutton-BrockJ. (1967): “Two cases of poisoning by contamination of nitrous oxide with higher oxides of nitrogen during anaesthesia”, Brit. J. Anaesth., 39, 388.
7.
DilworthN. M. (1973): “Hazard of O2 regulator outlet”, Anaesth. Intens. Care, 1, 549.
8.
EichhornJ. H., BancroftM. L., LaasbergL. H., DuMoulinG. C., and SaubermannA. H. (1977): “Contamination of medical gas and water pipelines in a new hospital building”, Anesthesiology, 46, 286.
9.
EpsteinR. M., RackowH., LeeA. S., and PapperE. M. (1962): “Prevention of accidental breathing of anoxic gas mixtures during anesthesia”, Anesthesiology, 23 (1), 1.
10.
FeeleyT. W., and Hedley-WhyteJ. (1976): “Bulk oxygen and nitrous oxide delivery systems: Designs and dangers”, Anesthesiology44 (1), 301.
11.
HoodJ. W. (1972): “Wrong gas connections”, Anaesth. Intens. Care, 1, 81.
12.
Hospital Technical Memorandum No. 22 (1972): Piped medical gases, medical compressed air and medical vacuum installations.Department of Health and Social Security, HMSO.London.
13.
IlsleyA. H., and CousinsM. J. (1978): “Study of the flow patterns of Cyprane Mk III vaporisers used for administering volatile anaesthetic agents.” In preparation.
14.
MazzeR. I. (1972): “Therapeutic misadventures with oxygen delivery systems: The need for continuous in-line oxygen monitors”, Anesth. Analg., 51 (5), 787.
15.
SpragueD. H., and ArcherG. W. (1975): “Intraoperative hypoxia from an erroneously filled liquid oxygen reservoir”, Anesthesiology, 42 (3), 306.
16.
Time Magazine, New York (15th August, 1977): Breath of death, p. 29.
17.
WylieW. D. (1975): “There, but for the grace of God …”, Ann. Coll. Surg. Engl, 56, 171.