Abstract
We analyze the pulsatile signal component of dynamic echo planar imaging data from the brain by modeling the dependence between local temporal and spatial signal variability. The resulting magnetic resonance advection imaging maps depict the location of major arteries. Color direction maps allow for visualization of the direction of blood vessels. The potential significance of magnetic resonance advection imaging maps is demonstrated on a functional magnetic resonance imaging data set of 19 healthy subjects. A comparison with the here introduced pulse coherence maps, in which the echo planar imaging signal is correlated with a cardiac pulse signal, shows that the magnetic resonance advection imaging approach results in a better spatial definition without the need for a pulse reference. In addition, it is shown that magnetic resonance advection imaging velocities can be estimates of pulse wave velocities if certain requirements are met, which are specified. Although for this application magnetic resonance advection imaging velocities are not quantitative estimates of pulse wave velocities, they clearly depict local pulsatile dynamics. Magnetic resonance advection imaging can be applied to existing dynamic echo planar imaging data sets with sufficient spatiotemporal resolution. It is discussed whether magnetic resonance advection imaging might have the potential to evolve into a biomarker for the health of the cerebrovascular system.
Keywords
Introduction
Propagating waves can reveal information about the medium they are traveling in. Wave characteristics, such as waveform, wavelength, velocity, and dispersion, often depend on the physical parameters of the medium. Cardiovascular pulse waves in the human body provide a natural physical disturbance to the vascular system whose effects have been utilized in clinical diagnostics for a long time. For example, pulse wave velocity (PWV) measurements in major arteries reveal information about arterial compliance.
1
Arterial compliance, the ratio of blood volume change to blood pressure change, is an important physiological factor by absorbing the impact of the cardiac pulse pressure waves and enabling steady blood flow throughout the whole cardiac cycle.2,3 The PWV is a reliable prognostic marker for cardiovascular morbidity and mortality in adult populations such as the elderly, patients with diabetes, and hypertension.
4
More important for the brain are the findings that aortic stiffness is associated with cerebral small-vessel disease in hypertensive patients,
5
that cognitive decline correlates with PWV,
6
and that neurodegenerative diseases such as Alzheimer’s disease are thought to be related to cerebrovascular dysregulation.
7
Furthermore, the understanding of emerging concepts such as pulse wave encephalopathy would be aided by novel diagnostic imaging methods of cerebral vasculature.8–12 However, there are only limited
In this article, we present phenomenological findings resulting from a spatiotemporal analysis of dynamic magnetic resonance imaging (MRI) data, as it is routinely acquired in functional imaging of the brain. Linear modeling of the dependence between local spatial and temporal signal variability results in a wave equation for unidirectional waves, a form of advection equation. The resulting imaging technique, which we term magnetic resonance advection imaging, or MRAI, provides velocity vector maps related to pulsatile dynamics obtained from dynamic MRI data. We aim at showing that MRAI has the potential to provide precisely localized information about pulsatile dynamics. This information might aid the interpretation of functional MRI and resting state brain network connectivity mapping, which is affected by vascular anatomy;20,21 for example, MRAI maps allow for the identification of local pulsatile components derived directly from echo planar imaging (EPI) data, without the need for co-registration. These components can induce long-range signal correlations reflecting vascular (in contrast to neuronal) connectivity. Although MRAI velocity maps are not yet quantitative maps of PWV, the results presented here might facilitate development of imaging biomarkers based on PWV in future.
In the Materials and methods section, MRAI and its application to functional EPI data are described. Also, models for spatial and temporal estimation errors are proposed. Finally, pulse coherence maps (PCMs) are introduced in order to demonstrate the pulsatile nature of MRAI maps. In the Results section, we provide evidence that MRAI maps pulsatile flow. 22 It is shown that inter-subject variability can be explained to a large degree, and potentially reduced in further applications, by the temporal estimation error model. The model is validated by numerical simulations. The phenomenological findings are discussed in the Discussion section with the help of further simulations, and possible future strategies to overcome present limitations are mentioned.
Materials and methods
MRAI model
Our goal is to estimate velocities of traveling waves in blood vessels from dynamic MRI data. The frequency-dependence of velocity, or dispersion, will be neglected, as will reflected pulses. In order to be able to generate spatial maps of velocity estimates, we define our model locally23–25 in a small spatial domain Ω in which the physical vessel and blood properties, and thus the pulse shape, do not change. In practice, this could mean that in a three-dimensional data set the model is defined for example for cubes of 3 × 3 × 3 voxels. The spatial velocity estimation maps, or MRAI maps, then consist of the combination over all those domains, defining the whole imaging volume. It is assumed that the dynamic MRI signal,
This equation simply linearly relates the temporal signal change to the spatial signal change, with the linearity coefficient
To understand the meaning of equation (2) we start with the one-dimensional advection equation, i.e.,
Equation (3) follows from the one-dimensional wave equation
This equation is fulfilled if either of the two factors vanishes. The first factor, a traveling wave in the positive x-direction for u > 0, yields equation (3), demonstrated in Figure 1(a). In three dimensions, equation (2) includes situations in which the wave velocity direction is not aligned with the spatial derivative of the signal; the dot product Traveling wave sampling scheme and multiple regression. (a) Schematics of a one-dimensional wave traveling along the positive x-direction with u > 0 and its spatiotemporal sampling by dynamic imaging. The sample points used for estimating the partial derivatives in the one-dimensional wave equation (3) shown at the top are labelled as ρ(x, t-Δt), etc. All these points lie in the same domain Ω. (This scheme emphasizes the waveform, which in real data is less pronounced.)
The local velocity
The terms Y and X1 to X3 are estimated from data by finite differencing as described below. The hat symbols indicate estimates rather than exact quantities. The constant term ‘1’ accounts for possible time-independent offsets in the data, for example gradients defined by brain anatomy. Additional nuisance variables are denoted by the symbol
The y and z derivatives are computed equivalently to the x derivative. This scheme has the advantage over more accurate higher order schemes that only nearest neighbor points of
Figure 1(b) shows an example for the regression applied to the data point marked with a red dot in Figure 2(c). Once model (2) has been fitted to each voxel time series of the dynamic MRI data via equation (6), MRAI maps can be obtained, showing either the magnitude (Figure 2(c)) or the direction of Magnetic Resonance Advection Imaging (MRAI). (a) Maximum intensity projection (MIP) of a time-of-flight MR angiography image of a healthy human brain at 7.0 T (Subject 16). (b) MIP of a PCM of the same subject, computed from a functional EPI scan. Bright pixels indicate a high maximum coherence between pulse oximetry time series and EPI signal. (c) MIP of an MRAI map. Bright pixels indicate high advection velocity coefficient estimates and the red dot marks a voxel that is used to demonstrate the regression approach in Figure 1(b). (d) Color-coded direction map displaying the directions of the estimated MRAI velocity vectors. Directions are indicated in the inset as green – anterior/posterior, red – right/left, blue – superior/inferior direction.
MRAI physiology
We assume that traveling waves described by the MRAI model above are related to blood volume waves. For example, for a blood vessel along the x-direction, equation (3) can describe the x-component of the pulsatile component of the blood flow velocity, and u is then defined as the PWV. The same advection equation also holds for blood pressure waves,
34
which, via the Windkessel model,
35
are a function of the integrated net flow into the vessel reservoir.
36
Since the local blood volume is related to local blood pressure via the compliance C as dV = C dp, the same advection equation applies to the pulsatile component of the blood volume (or vessel expansion) as well. The value of the velocity u is always the same, no matter whether the signal
If
As per this model, three vascular parameters determine the PWV: Vascular diameter d, wall thickness h, and the vessel’s elastic (Young’s) modulus or distensibility, E. The density of the blood,
Aliasing
A cerebral pulse wave with a velocity of 10 m/s has a wavelength of 10 m if the pulse frequency is 1 Hz. This is in stark contrast to typical EPI sampling rates, including the spatial sampling of 1.2 mm and the temporal sampling rate of 2 s of the data set analyzed for this work.
In the following, it is detailed how sampling affects the estimation of wave velocities via aliasing, and a correction factor is provided under the assumption that the pulse signals are periodic. In the Results section, it is shown that by using this correction factor the data can be unbiased to a certain degree, reducing inter-subject variability. The derivation of the correction, or bias factor, is given in Supplementary Information S1 and can be summarized as follows: the velocity is estimated from data by finite differencing applied to the wave equation (3). The resulting analytic relation shows that the
Equation (11) holds for our sinusoidal model with fixed T, but in practice the waveforms are more complex and T varies with the heartbeat, so the bias factor is an approximation.
Wavelength limits
The wavelength of a traveling pulse wave is much longer than a typical MRI spatial sampling interval and limits the maximum possible velocity estimates under given signal-to-noise ratio (SNR). In order not to compromise quantitative estimates of wave velocities in an unacceptable way, either spatial derivatives have to be estimated by other means than finite differencing, or the SNR has to increase considerably, or both. In Supplementary Information S2 an approximation for the maximum possible wave velocity is derived. It is validated in Figure 3(c) on numerical simulations.
Understanding estimation bias. (a) Regression between the average MRAI velocity coefficient estimates over the whole imaging volume for all 19 subjects against their average heart beat interval during the MRI scan. The two subjects with the lowest MRAI estimates, Subjects 1 and 2, as well as subject 16 from Figure 1, are labelled for comparison with the MRAI maps in Figure S1 and S2 and Figure 2, respectively. The regression is not significant (p > 0.05). (b) Regression between average MRAI velocity coefficient estimates and the magnitude of the aliasing factor (11) derived from a sinusoidal model of traveling pulse waves. The aliasing factor depends on the pulse period relative to the sampling rate of the data. It explains about half of the inter-subject variability and the regression is significant with p < 0.0005. The same subjects as in A are labelled here again. One can see that for example Subject 2, which had no useful MRAI image, has an aliasing factor of approximately zero, such that MRAI is not applicable to Subject 2 with the given sampling rate. (c) Estimated wave velocities from noisy simulations via Eq. (10). Shown are in dependence of given wave velocity u the expected estimated absolute velocity (black line), the estimates for an SNR = 32 (black asterisks), for an SNR = 64 (red asterisks), and for an SNR of 96 (green asterisks). In addition, the vertical dashed lines mark the respective expected maximum meaningful velocity. (Parameters: The SNR of 32 was determined from the data at the red dot of Figure 2(c) as the ratio of signal and background noise standard deviations. The expected wave velocities were corrected by a bias factor of -0.055 following from an average heart beat interval of T = 0.72 s from the same data set and the experimentally given temporal and spatial sampling rates of 2 s and 1.2 mm, respectively.)
Pulse Coherence Maps
Another concept defined at this place are pulse coherence maps (PCMs). They show the coherence between a pulse-related time series, for example obtained by a plethysmograph (pulse oximeter) from a finger, and each voxel time series of the dynamic MRI signal (Figure 2(b)). Even though the pulse frequency typically is higher than typical EPI sampling frequencies, PCMs clearly depict meaningful vascular signal components. Since the magnitude of the coherence is invariant under signal time shifts, PCMs do not require re-ordering or time-shifting the MRI data in order to take account for the unknown time delay between the pulse measured at the finger and in the brain.
22
The PCM is defined as
Here, coherence(.,.) denotes the normalized magnitude of the cross power spectrum density over its two arguments. It is a function of frequency, and the maximum is taken over a suitable frequency range. The frequency itself at which the maximum is attained is of no further interest as it normally approximates the average pulse frequency. The signal pulse(t) can be obtained from pulse oximetry data by resampling to the sampling time points of the dynamic MRI data. The signal
Natural stimulation functional magnetic resonance imaging data set
PCMs, MRAI velocity and direction maps were computed for all 19 subjects of a publicly available natural stimulation dynamic EPI data set obtained on a 7.0 T MRI scanner. 45 During scanning, subjects were listening to an audio version of a movie. This data set includes eight 15 min long segments for each subject of which the second one was used for analysis. The transversal slices covered most of frontal and occipital cortex and the regions in between. Data were sampled with a pulse repetition time (TR) of 2 s and an isotropic spatial resolution of 1.4 mm. The data set also contains time-of-flight angiography images of about the same coverage as the EPI data (Figure 2(a)), as well as pulse oximetry data.
For this study, all analysis was performed on MNI152-transformed, 46 geometric distortion, and motion corrected EPI data sets, with an isotropic spatial resolution of 1.2 mm. In order to test for possible effects of the transformation to MNI space, the analysis described below was also compared with selected raw data sets, with comparable results (not shown).
PCMs were computed from equation (12) with Matlab 2012b (The MathWorks, Inc., Natick, MA, USA), using the function cpsd, over all voxels inside the brain using masks. The maximum in equation (12) was computed over all but the lowest frequencies (cutoff = 0.0136 Hz), in order to exclude trend effects. The signal pulse(t) was obtained from the pulse oximetry data by spline interpolation (Matlab function interp1). The resulting maximum coherences were saved to a file for each of the subjects. PCMs (Figure 2(b)) were generated from these files by a maximum intensity projection (MIP) along the z-axis (head to foot) over slightly spatially smoothed maximum coherence values. For display, values inside the range [0.6, 1.0] were used in order to only show significant contributions.
MRAI velocity maps were computed with Matlab via equation (6): First, the data were high-pass filtered with a cutoff at 0.05 Hz (Matlab function idealfilter) in order to remove components caused by low-frequency signal components.47,48 Next, finite differences were computed following equation (8). The six motion parameter time series were used as additional nuisance regressors η in equation (6). Equation (6) was solved using a least-squares estimator (Matlab symbol “\”). The velocity estimate
MRAI direction maps were generated from the velocity component files by first assigning RGB values to the three velocity components: Red for ûx (brain right/left), green for ûy (brain anterior/posterior), and blue for ûz (brain superior/inferior) direction. Next, a MIP was generated by using only those voxels that found entry into the MRAI velocity MIPs as defined above. Finally, all RGB values were divided by the maximum RGB value in the MIP and the resulting map divided by a factor of 0.6 inorder to enhance colors.
Numerical simulations
In order to show performance and limitations of the MRAI fitting procedure with respect to both magnitude and direction of velocity vectors, numerical simulations of a traveling wave in a planar half-ring structure were performed. Although the MRAI model assumes physical homogeneity within the local domain Ω, this assumption is not always realized in experimental data and was taken into account with a spatial modulation profile across the ring structure, as described in the following.
Traveling waves were generated in Matlab for varying velocities as given in the leftmost column of Figure 4. Guided by possible multiband EPI acquisition parameters, the spatial resolution was set to Δx = Δy = 1 mm, the temporal resolution was Δt = 0.2 s, and 100 (space x) × 50 (space y) × 500 (time) data points were simulated. The center radius of the half-ring was r0 = 38 mm and the width was 19 mm. Traveling wave simulations were performed only in the region defined by this half-ring. Traveling waves were defined as traveling along the ring depending on the radius r2 = x2 + y2 in the following way: Local velocity coefficients Numerical simulations. Top row: Description of the simulated traveling wave on a curved path. The middle column shows the pulse wave velocities as gray values and the wave direction. The right column shows the color code for the velocity direction. The left column shows an amplitude profile reflecting narrow outer half-rings with increased/decreased signal amplitudes.

The modulation function m(x,y) consisted of a sinosoidal function of the radius only and was used to model the outer ring of increased/suppressed amplitudes. This modulation is visible in Figure 4, panel “Radial amplitude profile” and in the data snapshots. The period of the wave was T = 1.234 s and the noise ɛ(x,y) was normally distributed with mean zero and standard deviation of 0.01. With these parameters, the maximum velocity that could still be quantitatively estimated, would be about 25 cm/s. Modeled velocities were u = 1 cm/s, 10 cm/s, 1 m/s, and infinity, which corresponds to a standing wave (Figure 4, second row from bottom). In additon, MRAI maps were fitted to a model just consisting of the noise ɛ, without traveling waves (Figure 4, bottom row).
MRAI simulations were fitted in the same way as for the EPI data with the difference that the model was two-dimensional and no motion nuisance regressor was used.
Results
Natural stimulation fMRI data set
Pulse coherence, MRAI velocity, and MRAI direction maps of natural stimulation data acquired on 19 subjects on a 7.0 T MRI scanner 45 are provided in Figure 2 and Supplementary Figures 1 to 19 along with MR angiography images. Figure 2(a) shows an angiogram with annotations of major cerebral arteries and the superior sagittal sinus. Figure 2(b) shows a PCM MIP. Figure 2(c) shows a MIP of the MRAI velocities, and Figure 2(d) shows a color-coded MIP of the direction of the MRAI velocities.
PCMs depict regions of high coherence between the pulse oximetry data and the EPI signal in the brain. The outline of some of the major arteries and the sagittal sinus, as seen in the angiography images, is clearly visible in the majority of subjects.
MRAI velocity maps often look similar to the PCMs but seem to be better defined with respect to the location of vessels. In particular, in some subjects parts of the following structures are visible: anterior cerebral artery, middle cerebral artery (including angular branch), posterior cerebral artery (including calcarine branch), and the superior sagittal sinus as a prominent venous structure. There is a large inter-subject variability which will be partially resolved below. Quantitatively, estimated MRAI velocities are significantly smaller than expected PWVs and even blood flow velocities. 49 For example, in Subject 16, Figure 2(b), the maximum velocity estimate is û = 3.5 mm/s, corresponding, after bias correction with equation (11), to a corrected velocity of 6.3 cm/s. MRAI direction maps show velocity vector directions in all three spatial directions. Most often, but not always, the estimated velocity direction corresponds to the vessel orientation as seen in the MR angiogram.
Aliasing correction to reduce inter-subject variability
For the specific data sets used here with a sampling interval of Δt = 2 s, following equation (S6), heart beat intervals of T = 4/n s = 4.000, 2.000 1.333, 1.000, 0.800, 0.667, 0.571, 0.500,… s would be expected to be unsuitable for MRAI. Now, the two cases with the smallest average velocity estimates, Subjects 1 and 2 (see Supplementary Figures 1 and 2), have an average period T of 0.816 and 0.800 s, respectively, which are very close or identical to the forbidden value of 0.800 s. Following is Subject 19 with a period of T = 0.551, still close to the forbidden value of 0.571 s.
In more general terms, a regression analysis between MRAI velocity estimates and the magnitude of the bias term from equation (11) for all of the 19 subjects yields a correlation coefficient R2 = 0.52 (p = 0.0005; Figure 3(b)). Figure 3(a) shows that the average pulse period T itself does not account for the quality of the MRAI maps. These results indicate that the sinusoidal pulse wave model can be used to optimize sampling dependent upon heart rate, or to obtain calibration factors for MRAI.
Numerical simulations
Simulation results for planar traveling waves with small additive noise and spatial signal modulation in curved vessels are shown in Figure 4. The following can be observed: (i) For a small velocity magnitude of 1 cm/s, the average estimate corresponds well to the expected average velocity given by equation (11). The modulation in the outer ring affects the estimates only slightly. The direction estimates correspond to the true velocity direction. (ii) For a larger velocity of u = 10 cm/s and same wave period, a similar result holds, although in this case the spatial extension of the wave, or wavelength, is already comparable to the whole ring. (Note that by keeping the wave period fixed, increasing the velocity increases the wavelength accordingly). (iii) For an even higher velocity of u = 1 m/s, the spatial signal change in the wave snapshot is no longer discernible, and the average MRAI velocity becomes strongly underestimated. Now, there is a clear difference between the spatially homogeneous component of the simulation and the modulated component in the outer ring, in which velocity estimates are larger than in the homogeneous component (but still underestimated; blue arrow). (iv) For an infinitely large wavelength (the first term in the argument of the sine in model (13) vanishes), or a standing pulse wave, MRAI velocity estimates are even more reduced. In this case, there is strictly no longer a traveling wave with a wave direction defined. However, the MRAI direction map still shows an apparent wave direction in the signal modulated component (green arrows). As discussed below, this is an artifact resulting from a misspecification of the MRAI model within the modulated signal component.
Discussion
MRAI
In MRAI, cerebrovascular pulses are modeled
Bias by aliasing and noise
The PWV estimates
Bias by model misspecification
An important model assumption contained in the expression (1) is that the velocity
Bias by model misspecification interfering with noise
Signal modulation can also interfere with the aforementioned noise bias. For example, if the noise is constant but the signal modulated in space, then the SNR varies over space. Figure 4, middle column, fourth row, shows such a case; within the modulated component of the simulation, the velocity estimates are larger (blue arrow, bright gray values) than in the unmodulated component of the simulation. Here, the magnitude of the modeled velocity was 1 m/s, and noise limits estimation accuracy. For smaller velocities of 10 cm/s and 1 cm/s (in the two panels above), the estimates within the modulated component do not differ significantly from the estimates in the unmodulated component.
A more subtle effect of noise in spatially modulated signals is that the estimator for
Other forms of errors
In particular, for fast wave velocities, data acquisition timing conditions such as slice-timing or the time necessary to fill lines in k-space in the MRI data set need to be taken into account, which we have not done here. It would require modification of pulse sequences and/or corrections in how the derivatives are estimated. Due to the nature of the inverse problem of estimating PWV, it is not guaranteed that we have found all possible confounds that potentially can bias the result. There might be additional biases caused by model assumption violation, for example the existence of multiple solutions that might occur for a superposition of traveling waves with two different velocities. In simulations with curved structures, we have found, for example, incorrect results caused by apparent shear flows that forced the regression problem into unexpected solutions. These solutions could be prevented by adding a miniscule amount of noise to the data, however, and do not seem to be of any relevance for realistic applications to dynamic MRI data. There might also be other biases resulting from motion, trends in the data, atypical noise sources, etc. More research is needed to evaluate whether these are relevant for the estimation of pulse related velocities in cerebrovascular data.
Pulse Coherence Maps
The PCMs presented here clearly show the effect of pulsatile flow, too. The important difference between PCMs and MRAI is that the former require a reference pulse waveform. In MRAI,
Clinical relevance
It is difficult to state any clinical relevance of MRAI at this point, for two reasons: MRAI velocities are not yet quantitative estimates of PWVs, and we have only used healthy control data. To determine whether MRAI velocities are a proxy for PWVs, empirical studies on subjects with cerebrovascular disease or comparison with a gold standard for PWV measurement in the brain would be necessary. In this regard, MRAI has the potential to provide valuable local measurements: For example, increased PWV causes an earlier wave reflection and increased systolic pulse in older adults,
9
as has been found in studies of the aortic PWV,
51
extending pulsations into smaller vessels and increasing stress on vessel walls, causing vessel degeneration. Furthermore, PWV increases in a much more characteristic way with age then brachial pulse pressure.
51
Analogously, an increased
Conclusions
Two concepts for imaging pulsatile cerebrovascular dynamics have been introduced: Pulse coherence maps and MR advection imaging. Whereas the former concept requires a reference pulse signal, the latter does not. PCM and MRAI images appear to be similar, but MRAI seems to be better defined with respect to vascular anatomy. In addition, MRAI provides regional estimates of wave velocities.
For the chosen temporal and spatial resolution, the observed MRAI maps are mainly affected by pulsatile signal components, as has been demonstrated with comparison to PCMs. At this point, further research is needed to identify all possible biophysical origins of vascular signal change; the advection model was intended to model traveling pulse waves, but blood flow, oxygenation changes, or other yet unknown factors might affect the signal as well. Nevertheless, due to its sensitivity to pulsatile components of the signal and due to dramatic advances in dynamic MRI data acquisition, MRAI might have future potential to contribute to the modeling of the cerebrovascular system and to serve as a biomarker for cerebrovascular disease. Finally, it should be noted that the methods described herein are quite general, and could in principle be applied to pulsatile dynamics across a wide range of dynamic imaging applications in medicine and other fields.
Funding
HUV acknowledges support by the Nancy M. and Samuel C. Fleming Research Scholar Award in Intercampus Collaborations, Cornell University.
Footnotes
Acknowledgements
The authors acknowledge useful hints for possible improvements towards quantification from the reviewers.
Declaration of conflicting interests
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Authors’ contributions
HUV designed and performed the study and wrote the manuscript. JPD contributed to physics discussions and wrote the manuscript. KT contributed to mathematics discussions and wrote the manuscript. NDS provided clinical input and wrote the manuscript. DJB contributed to physics discussions and wrote the manuscript.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.
