Abstract
PURPOSE:
Based on a retrospective analysis, this study aims to develop a simple index for validity of the evaluation point for the dosimetric verification of intensity-modulated radiation therapy (IMRT).
METHODS:
The results for the dosimetric verifications of a total of 69 IMRT plans were analyzed in this study. A Farmer-type ion chamber was used as a dose detector, and a solid water-equivalent phantom was used. Index values were obtained by dividing the difference between the maximum and minimum dosages by the mean dosage of the 69 plans, and the values were classified into five groups with index value <4, 4–8, 8–12, 12–16, and >16. A
RESULTS:
We found that there was no significant difference between the groups with index value <4 and 4–8 (
CONCLUSIONS:
The results obtained in this study may prove to be useful for accurate dosimetric verifications of IMRTs when ion chambers are used.