Abstract
BACKGROUND/OBJECTIVE:
There are two routine pulse-sequences for single voxel spectroscopy (SVS), point resolved spectroscopy (PRESS) and stimulated echo acquisition mode (STEAM). Although STEAM has several advantages in comparison to PRESS, signal/noise ratio (SNR) superiority of PRESS makes it the first choice for SVS. Application of fast Padé transform (FPT) instead of Fast Furrier transform (FFT) might increase the SNR of the signal produced by STEAM pulse-sequence and therefore allows the benefits of its advantages. We aimed to evaluate and compare the noise root mean square (RMS) and SNR provided by STEAM pulse-sequence using both FPT and FFT.
MATERIALS AND METHOD:
A gelatin-based phantom was constructed in a 19-cm acrylic cylinder. The phantom had two normal/tumoral parts. The SVS was performed using a 3T MRI scanner. STEAM pulse-sequence were used with the following parameters: TR
RESULTS:
According to the results, noise RMS of spectra provided by FPT were decreased between 3619.01–14252.94% in comparison to FFT (
DISCUSSION:
The difference of noise RMS’s provided by FPT are thousands percent less than FFT. This enormous decrease in noise provides a good increase of SNR. While the range of Cr1 and Cho SNR by FFT are between 41.55–120.32 the range of SNRs of these peaks provided by FPT are between 1719.99–9744.79, which implies a significant difference between the efficiency of FPT and FFT.
CONCLUSION:
This study showed that application of FPT in comparison to FFT can increase the spectra SNR and so that its usage can be helpful during the application of STEAM pulse-sequence which results in lower SNR in comparison to PRESS pulse-sequence. Thus, we should make use of the advantages of STEAM pulse-sequence.