Abstract
Background
Velocity-encoded magnetic resonance imaging (VENC-MRI) is a commonly used technique in cardiac examinations. This technique utilizes the phase shift properties of protons moving along a magnetic field gradient. VENC-MRI offers a unique way of measuring the severity of valve regurgitation by directly quantifying the regurgitation flow volume.
Purpose
To compare flow analysis results of different software programs and to assess the effect of background correction in sample patient cases.
Material and Methods
A phantom was built out of Polymethyl methacrylate (PMMA) which provides tubes of different diameters. These tubes can be connected to an external water circuit to generate a water flow inside the tubes. Expected absolute flow quantities inside the tubes were determined from preset tube- and flow-parameters. Different flow conditions were measured with a VENC-MRI sequence and the images evaluated using different software packages. In a second step six randomly selected patients showing different degrees of aortic insufficiency were evaluated in clinical terms.
Results
The contour propagation algorithms used in the software packages performed differently even on static phantom geometry. In terms of clinical evaluation the software packages performed similarly. Enabling background correction or leaving out manual correction of propagated contours changed results for severity of aortic insufficiency.
Conclusion
Turning on background correction and manual correction of propagated contours in MRI flow volume measurements is strongly recommended.
Keywords
Introduction
Velocity-encoded magnetic resonance imaging (VENC-MRI) is a commonly used technique in cardiac examinations (1,2). There is also growing interest for using this technique in brain examinations (3,4).
VENC-MRI utilizes the phase properties of protons moving parallel to a magnetic field gradient. These protons undergo a phase shift which is proportional to their velocity. Subtracting a reference scan from a velocity encoded scan results in phase contrast velocity maps. Clinical applications of VENC-MRI include measurement of valve regurgitation severity (5–7) or detection of hemodynamic significance of stenosis (8).
For patients with valve regurgitation VENC-MRI is a unique technique compared to other imaging modalities as it can directly quantify the regurgitation in mL/min (9). A device used to directly quantify a physical property needs quality assurance (QA). Summers et al. (10) presented a phantom for flow measurement and assessed it in a multisite trial. Their phantom consists of straight, stenosed, and U-bend flow channels embedded in polymer layers. These static agarose layers can be used for background offset correction.
Inter-study reproducibility of VENC-MRI was assessed and the variability was found to be very low (11). One of the major reasons for errors in velocity encoded measurements is the existence of field inhomogeneities inside the scanner. These lead to eddy current effects which will in turn introduce spatially dependent phase distortions (12). These distortions or phase offsets can be corrected by analyzing areas with static tissue to derive a correction function for the offset error (12,13). Phase errors caused by concomitant fields (14) are corrected during postprocessing as implemented in the scanner software.
To analyze flow through a specific region of VENC-MRI images a region of interest has to be delineated. In case of patients with aortic insufficiency images of a two-dimensional (2D) slice perpendicular to the aorta are acquired at multiple time frames throughout the R-R interval of the heart cycle. As the cross-section and the position of the aorta changes over time the region of interest (ROI) has to be adjusted for all time frames. Van der Geest et al. (15) presented an automatic contour propagation algorithm where only the center of the aorta on one time-frame needs to be selected by the user. Automatic propagation could be done in about 6 s compared to 10 min of manual contouring.
Inter- and intra-observer variabilities for manual and automatic detection of the aortic contour was shown to be low demonstrating that the automatically detected contours detection yielded reproducible results (15). Misalignments of contours by errors in propagation algorithms present an additional source of errors for VENC-MRI measurements.
The aim of this study was to compare flow analysis results of different software programs and to assess the effect of background correction and contour misalignment for sample patient cases. Comparison results should show the impact of the assessed effects and from that give a guideline for necessary steps in the physician’s workflow.
Material and Methods
Phantom
The phantom consists of a cylinder out of Polymethylmethacrylate (PMMA) with a diameter of 300 mm. By filling the phantom with distilled water a signal for phase offset calculations is provided. Six tubes showing different flow configurations (e.g. constant tube diameter or varying tube diameter to simulate different degrees of stenosis) were placed inside the phantom. The diameters of the inner tubes were 24 mm and 10 mm with several possible stenosis lengths. A seventh tube inside the phantom allows for inserting smaller tubes with customized flow configurations. Fig. 1 shows a three-dimensional (3D) model of the phantom.
3D model of the phantom.
A water circuit was designed to enable continuous measurements of different flow rates (Fig. 2). To allow for simple changes and direct control of flow rates the pump was placed outside of the MRI room. A frequency converter was used to change flow rates of the water pump. The flow rates were measured by rotameters directly behind the pump for comparison to the MRI measurements.
Flow circuit set-up.
Expected mean flow velocity inside the tube was calculated with following formula:
Software
Four different commercially available software packages and manual postprocessing done in MATLAB (The MathWorks Inc., Natick, MA, USA) were compared. The software packages used were Siemens Argus Flow (Siemens Healthcare, Erlangen, Germany (16)), circle CVI42 (Circle Cardiovascular Imaging Inc., Calgary, Canada (17)), Medviso Segment (Medviso AB, Lund, Sweden (18)), and Medis Q-Flow (Medis Medical Imaging Systems BV, Leiden, The Netherlands (19)).
Patients
Six patients were selected to cover regurgitation fraction in the range of 0–40%. The age of selected patients was in the range of 21–72 years (5 women, 1 man).
Image acquisition
Images were acquired on an Avanto 1.5T and on a Skyra 3T (Siemens Healthcare) using Siemens’ standard VENC encoded flow sequence “FLASH”. For the phantom measurements an ECG curve was simulated and used to acquire eight time-frames with a TR of 117 ms, a TE of 3.34 ms, a flip angle of 20 degrees, a slice thickness of 6 mm, and an in-plane resolution of 0.9 mm. Patient’s R-R intervals were acquired in 30 time-frames using the same sequence with enabled ECG triggering, a TR of 39.76 ms, a TE of 2.68 ms, a flip angle of 20 degrees, a slice thickness of 6 mm, and an in-plane resolution of 1.77 mm.
Software operating procedure
The operating procedure for the phantom was to delineate the tube on the first time-frame, then run the contour propagation algorithm to create delineations on all succeeding time-frames. No manual correction was applied afterwards.
For patient cases, contours were manually corrected if needed. Resulting forward and backward flow volumes were collected for original automatic contours and for manually corrected ones to inspect differences. Additionally also background corrected values were calculated for both cases. All software packages support “static tissue background correction” so this correction was applied in all cases. A region inside the back muscles of each patient was selected as static-tissue region. As there is no contour propagation involved in the background correction calculation the same correction values were applied in all software packages.
Manual postprocessing in MATLAB was done for the phantom measurements by selecting the ROI pixel-by-pixel on magnitude images.
Results
Phantom
Mean flow velocity and area results for phantom measurements.
Software
Available features in software programs used for this study.
Available in standard viewer of Siemens system but not in the Argus window.
No stand-alone, only available for a Siemens Multimodality Workplace or for a Siemens MR.
BG, background; ROI, region of interest.
Patients
Regurgitant fraction in % representing the severity of aortic insufficiency.
BGC, with background correction.
Highest difference for enabling background correction of 7.2% was found using Q-Flow for patient P1. Mean velocity offset was −2.5 cm/s with a maximum velocity of 96 cm/s in this case.
Discussion
In clinical day-to-day routine of analyzing aortic flow, physicians contour the aorta on one time-frame where the aorta is clearly visible and let the software propagate this first contour onto all cardiac phases. In a second step, contours are typically corrected manually. For clinical evaluation the quota of backward flow volume to forward flow volume (regurgitant fraction) is analyzed. According to this clinical procedure six patient cases were compared by first delineating the aorta where it showed well visible borders. Then the contour propagation was run and afterwards the contours were manually corrected if needed.
For the phantom measurements CVI42 and Segment showed best agreement in terms of detected area and mean velocity inside the tubes. Using MATLAB to manually mask the tubes in the images yielded similar results. No manual correction was applied afterwards as the goal was to test the detection of a simple clearly delimited circular geometry.
Q-Flow showed the worst results for the phantom measurements but this could be due to some part of the algorithm which does not affect clinical evaluation. The exact functional principle of the algorithm is unknown at this point but it is important to note that there are such large differences in contour propagation. In response to an inquiry by email Medis (Medis Medical Imaging Systems BV, Leiden, The Netherlands (19)) pointed out that the contour detection was revised in their new Version 5.6 which our institute was not yet upgraded to.
Comparison of patient evaluation showed that the effect of background correction is non-neglectable. Gatehouse et al. (20) showed that a velocity offset correction of 2.4 cm/s could result in up to a 10% difference in regurgitant fraction. They also proposed that the maximum offset which does not need correction is 0.6 cm/s. Mean offset error was 2.7% for their multi-site phantom based study. These results are in agreement with the results presented in this paper where a mean velocity offset of −2.5 cm/s showed a difference in regurgitant fraction of 7.2%.
One limitation of the study presented here was the estimation of phase offset due to magnetic field inhomogeneities by deriving a linear correction function from areas containing static tissue (12–14). This first approximation of background offset was chosen as this type of correction is supported by all compared software packages. To use phantom calibration measurements is a more precise method for background correction (21) but doubles measurement time and is not done in the clinical routine of our institute (see Table 1 for software packages which provide phantom offset correction).
Giese et al. (22) presented an alternative approach to use magnetic field monitoring to calculate background velocity offset corrections. They showed that echo times can be optimized to minimize background offset and by additionally using the calculated offsets, the errors could be decreased to less than 0.5%. This is a promising method for background correction as it is very accurate and does not lengthen measurement time.
A second limitation is the size of the patient set. The purpose of this study was not to show statistically significant differences in contour propagation but to confirm that manual contour correction should be done by default and to show that background offset correction has an impact on clinical evaluation of patients. Ignoring errors of automatically propagated contours lead to differences in regurgitant fraction of the same dimension as the ones resulting from background offset. This could mislead the clinical evaluation of the severity of aortic insufficiency.
The software package Segment being freely available for research purposes showed reproducible results and provides the highest number of features for flow evaluation. Segment provided all commonly used background correction methods.
In conclusion, contour propagation algorithms showed differences between software programs. Even for a simple static phantom geometry differences in contour detection were seen. For patient cases the contour propagations varied between patients and between programs. Automatic contour propagation has to be checked and corrected manually if there is a visible discrepancy. In terms of clinical evaluation of regurgitant fraction the different software programs performed similarly.
Enabling background correction or leaving out manual correction of propagated contours lead to non-neglectable changes in regurgitant fraction. Therefore turning on background correction and checking automatically propagated contours for validity is recommended for all MRI flow volume measurements.