Abstract
Background
Light exposure is an important driver and modulator of human physiology, behavior and overall health, including the biological clock, sleep-wake cycles, mood and alertness. Light can also be used as a directed intervention, e.g., in the form of light therapy in seasonal affective disorder (SAD), jetlag prevention and treatment, or to treat circadian disorders. Recently, a system of quantities and units related to the physiological effects of light was standardized by the International Commission on Illumination (CIE S 026/E:2018). At the same time, biometric monitoring technologies (BioMeTs) to capture personalized light exposure were developed. However, because there are currently no standard approaches to evaluate the digital dosimeters, the need to provide a firm framework for the characterization, calibration, and reporting for these digital sensors is urgent.
Objective
This article provides such a framework by applying the principles of verification, analytic validation and clinical validation (V3) as a state-of-the-art approach for tools and standards in digital medicine to light dosimetry.
Results
This article describes opportunities for the use of digital dosimeters for basic research, for monitoring light exposure, and for measuring adherence in both clinical and non-clinical populations to light-based interventions in clinical trials.
Keywords
Introduction
The importance of light for human health and well-being
Exposure to light has a powerful impact on human health and well-being.1,2 This impact is mediated by a range of physiological responses to ocular light exposure. By illuminating the world around us, light enables us to see and perceive the world, contributing to visual performance, inducing visual experiences and affecting visual comfort. Beyond vision, light is crucial for everyday function: It is the main time cue for the circadian system, which for example regulates the phase and amplitude of an individual's daily sleep-wake pattern 3 but also, basic physiological functions such as immune function 4 and metabolism. 5 In addition, a well-functioning circadian system is important for healthy sleep and optimal functioning during wakefulness. Disturbances in sleep-wake patterns due to the absence of a strong time cue or an ill-timed signal for the circadian system can result in a lack of energy, cognitive deficits, reduced self-control and a negative mood.6–8 Light can also induce more instantaneous changes in alertness, mood, and performance,9–12 independently of the crucial role of light in sleep and circadian functioning. Through these acute effects, light can – depending on its strength and timing – benefit or challenge our daytime functioning and overall physiology. Recently, consensus-based recommendations for healthy light exposure patterns were put forward by an international group of experts 13 – further highlighting the currently under-recognized potential of light for human health. Building regulations and recommendations have been developed to address specifically the requirements of illumination vis-à-vis the physiological effects of light14–16 (reviewed by Ref. 17 ).
Light exposure needs to be measured in a personalized fashion
With light playing such an important role in controlling and modifying various aspects of our physiology, behavior, and long-term health, the measurement of light is of great importance. The measurement of light – called optical radiation metrology – is a very mature field, with many commercially available high-quality instruments for stationary measurements, such as those of a specific light source. However, humans are not stationary, moving around in and between different environments that are illuminated by diverse light sources, including electric light, daylight and self-emitting displays such as computer monitors or smartphones, which are themselves not constant due to weather conditions and daylight availability.18,19 Rather than measuring light in a location-specific or source-centric view, there is therefore the need to characterize light exposure in an personalized fashion. 20 Importantly, such characterization is not limited to the physiological “non-visual” responses to light, but can equally include the characterization of the “visual diet”, i.e., which types of colors and illuminances people are exposed to. 21
Various research groups and commercial manufacturers have developed measurement or sensor solutions that are person-based, as they are worn on different parts of the body (e.g., as a brooch, wrist-worn watch-like device, on a pair of glasses, etc.), and allow for prolonged monitoring of light exposure in free-living conditions.21–30
Critical factors influencing the measurements include the position of the sensor (and e.g., the possibility of the sensor being covered by clothing such as sleeves or scarves), the vulnerability of the sensor (e.g., waterproof grading, ruggedness, position on the body and variability of that position31–34), battery life, primary function of the sensor (e.g., is it a primary actigraphy monitor, optical monitor, or a light sensor designed to capture different dimensions of environmental light exposure), and acquisition-related parameters such as the recording interval. Systematic in-laboratory comparisons of the optical performance of light-logging sensors show large differences in their ability to capture biologically relevant light in commonly used sensors.25,34–40
A recent comprehensive review on light-dosimetry studies showed that in the published literature, there is a large variability of parameters used in these studies. 20 The authors found a total of 25 unique device types used in the literature. In more than 70% of the surveyed studies, there was no mention of the dosimeter's calibration. Device placement was found to vary widely between studies, with the wrist location being the most commonly used one (64% of the surveyed studies). The recording interval also varied, with 1-minute intervals being the most commonly used one (49% of the surveyed studies). Given this diversity, it is clear that a framework is needed to understand the choice of different parameters systematically.
Typical workflow for measuring personalized light exposure
Figure 1 visualizes the general workflow of personalized light-dosimetry measurements. The world contains scenes that are illuminated by different light sources, including daylight, electric light of various types, and mixtures thereof. Scenes contain materials that reflect light in potentially spectrally biased ways. Consequently, the light reaching the eye – ocular light exposure – is a complex combination of light sources, illumination geometry, and materials, all of which may be changing over time. Ocular light exposure itself is modified by dynamic parameters, such as head and eye movements, and individual-level parameters, such as facial features.

Overview of the light-dosimetry and light-logging pipeline.
The quantity that drives the physiological effects of light – and the associated health outcomes – is the irradiance reaching the retina in the back of the eye, after passing through the pupil and being filtered by the ocular media. As retinal irradiance is not measurable by external tools – only the retina can “measure” it 41 – proxy measurements using irradiance sensors in, near or outside of the corneal plane are used to approximate retinal irradiance. Most commonly, wrist-worn devices with build-in light-logging facilities are used (64% of the surveyed studies in Ref. 20 ). When worn continuously, these measurement sensors yield time series of personalized light exposure. For research purposes, these are then typically summarized using level- (e.g., mean or median) or time-based (e.g., time above threshold) metrics and related to health-related outcomes by way of statistical association.
Of course, at the start of each investigation using personalized light exposure measurements is a research question. Different research questions will prioritize different methodological choices and outcomes: Some research questions may not require a measurement device to be accurate. In the following, we approach the question of light dosimetry from a device-dependent perspective, rather than considering the many idiosyncrasies that could arise in different research questions.
Applications of measuring personalized light exposure in medicine and health
There are broadly four distinct modes of using light dosimetry for investigating health-related outcomes:
Monitoring light exposure in observational studies
Light exposure patterns among free-living subjects have been captured in a number of studies. This includes research primarily describing and characterizing the patterns of light exposure across seasons and in different populations,26,42–47 as well as studies associating light exposure and health-related outcomes.48–58 The biological rationale for hypothesized associations between light exposure and health-related outcomes is clearly established through in-laboratory findings in humans and animals.2,13
Monitoring light exposure as an outcome or mediator in clinical trials
In some cases, light exposure can represent the primary or secondary outcome of a specific clinical trial, such as in studies attempting to modify lifestyle to alter light exposure and visual behavior in myopia control (e.g., Ref. 30 ), or studies aimed to optimize light exposure in order to support sleep and/or daytime functioning (e.g., Ref. 59 ). In these studies, personalized light exposure is the study outcome, and therefore measurements need to be sensitive enough to capture change in light patterns in response to the trial intervention. Similarly, studies that focus on lifestyle interventions to improve sleep might affect light exposure (e.g., an increase in physical activity outdoors might be the behavioral intervention), and the potential increase in light exposure may affect sleep. Measuring the direct and indirect effects of an intervention, in turn, informs our causal models of disease, and enables us to maximize impact in future study designs.
Monitoring light exposure to confirm compliance in clinical trials and in-laboratory studies
Light is used as an intervention in a range of disorders and syndromes, including depression, 60 postpartum depression, 61 cancer,62–64 dementia 65 and delayed phase sleep disorder. 66 Typically, these interventions involve the application and exposure to bright light sources in an ambulatory or stationary setting. For participants and patients who can freely move around, light dosimetry can therefore be used for confirming the actual effective dose (and thereby adherence to the intervention). Similarly, light exposure can be measured prior to laboratory sessions to estimate the effective dose in in-laboratory interventional studies. 67
Monitoring light exposure for diagnostic purposes
There are wide reaching consequences of environmental light exposure, 68 some of which have been the focus of large-scale epidemiological studies where light exposure is measured using either stationary sensors (e.g., to assess light at night in a bedroom) or at an ecological level (e.g., using satellite data for certain geographical location, often averaged across time). These studies have provided some evidence that exposure to ambient light at night is associated with an increased risk of diabetes, 69 breast cancer,70–72 depression 73 and obesity.50,74–77 Effect sizes are generally small, and replication of longitudinal studies using incident disease outcomes is scarce. In addition, biological research clearly shows that the neural responses caused by light exposure are tracked and summed over time and that physiological responses depend on prior light history,78–81 so non-personalized light exposure measurements, especially over a short timeframe, are likely to misestimate the true impact of light exposure on long-term health.
The V3 framework: verification, analytical validation and clinical validation
The main goal of this article is to outline the steps required to advance personalized light exposure assessment to become fit-for-purpose as a medical intervention and health-related outcome and predictor in (digital) clinical trials. To do so, we have adopted the V3 framework (Box 1) which was developed to enhance reliable and robust usage of biometric monitoring technologies (or BioMeTs) in clinical research. 82 This framework includes verification (Figure 2A), analytical validation (Figure 2B) and clinical validation, which we apply to the use of light dosimeters (Figure 2).
V3 framework
The V3 evaluation framework 82 is a three-component process intended to provide guidance on evaluating biometric monitoring technologies (BioMeTs) as fit-for-purpose. BioMeTs are connected digital medicine products that process data captured by mobile sensors using algorithms to generate measures of behavioral and/or physiological function.
The framework includes (1) verification, (2) analytical validation, and (3) clinical validation, described in turn below. The V3 evaluation framework was developed by the Digital Medicine Society to synthesize the various definitions and standards that had already been in use within engineering, regulatory, and clinical fields.
Put simply, the V3 process aims to answer the following questions in turn:
To what extent does the output of the sensor measure the physical concept of interest within acceptable performance criteria (i.e., is the sensor contained within the device/product adequately capturing light exposure)? To what extent does the output of the algorithm capture the relevant behavioral or physiological phenomenon it claims to measure (e.g., is the output data an accurate measurement of physiologically relevant light exposure)? Is the behavioral or physiological phenomenon clinically relevant for the population of interest and context of use (e.g., does circadian phase estimated with light sensor data predict a relevant downstream health consequence)?

Mapping the V3 model onto light dosimetry and light logging.
Verification of light dosimeters
The goal of verification is to determine if the light sensor is accurate with respect to a reference standard, precise (intra-sensor assessment over short time periods), consistent (intra-sensor assessment over longer time periods), and uniform (inter-sensor assessment). In the case of light sensors, the International Commission on Illumination (CIE), the standards body for light-related quantities, has yet to release a standard characterizing the properties of light sensors for personalized light exposure. For illuminance meters, the international standard ISO/CIE 19476:2014 – Characterization of the performance of illuminance meters and luminance meters
83
considers a range of properties when determining the performance of illuminance meters. Though this standard focuses specifically on measurements of photopic illuminance, its general principles of calibration are useful and can be applied to light sensors capturing biologically relevant light exposure (Figure 2A). We consider here the most important points:
Metrology for the visual and non-visual responses of light When discussing light, it is important to understand that the measurement and characterization of light is intimately linked to human physiology and behavior. Illuminance [lux] and luminance [cd/m2] are photometric quantities that imply a human observer, as they represented the spectral irradiance or radiance as weighted by the photopic luminous efficiency function V(λ). V(λ) was derived in 1924 by the International Commission on Illumination (CIE; Commission Internationale de l’Éclairage) from psychophysical measurements in a group of human observers, and generally V(λ) can be modeled as a weighted sum of the long-wavelength-sensitive (L) and middle-wavelength-sensitive (M) cone spectral sensitivities. In 1931, the CIE developed a system of colorimetry to describe color objectively, resulting in the CIE 1931 XYZ system (in which V(λ) is included as the ȳ(λ) function). In 2018, a novel system of metrology was developed by the CIE to quantify the activation of all human photoreceptors with specific relevance to the physiological and behavioral effects of light, incorporating also the melanopsin-containing ipRGCs in addition to the long-wavelength-sensitive (L), medium-wavelength-sensitive (M) and short-wavelength-sensitive (S) cones and the rods. This system proposes a set of five spectral sensitivities to calculate the α-opic (ir)radiance, which is the (ir)radiance for a specific retinal photoreceptor class (S cone, M cone, L cone, rhodopsin and melanopsin-encoded photoreception of ipRGCs). It is possible to convert the α-opic values into a colorimetric system like the previously described CIE 1931 XYZ system.
Taken together, the verification process and the data generated in this process is often opaque in the context of BioMeTs, and certainly so for wearable personal light sensors. This article provides recommendations for the evaluation of light sensors, especially when employed in clinical research settings. While it may not be straightforward to perform these characterizations, the V3 framework and its explicit call upon verification of BioMeTs for clinical research will support the development of a consensus and/or community-driven standards and guidelines.
Analytical validation of light dosimeters
The analytical validation of light dosimeters refers to the validation of light measurements under well-controlled lighting scenarios made with the dosimeters (Figure 2B). This is most convenient under laboratory conditions, where lighting conditions can be held constant and accurately calibrated using high-specification research-grade stationary spectrometers, hyper-spectral cameras or luminance cameras. If a BioMeTs under investigation generates light measurements that inaccurately represents light exposure under these parametric conditions, it is very unlikely that it can also support measurements in the context of clinical trials in real-world settings. Depending on the types of sensor channels contained in a dosimeter, there are obvious ways to check the (ecological) validity of the measurements. For example, if a dosimeter has a UV channel, then under typical indoors without any daylight condition, the UV irradiance should be minimal. However, in outdoor light conditions, measurements might be heavily affected. A relatively weak but occasionally useful form of analytical validation is the cross-calibration of BioMeTs to be used in a study. This could involve subjecting the light dosimeters to the same set of standard conditions – e.g., an overcast sky 84 – and then generating a correction factor to at least make devices used within a study to yield comparable data.
Clinical validation of light dosimeters
The clinical validation of light dosimeters demonstrates their usability in describing, predicting or explaining health or clinically relevant outcomes (Figure 2C). As described above, there are various studies that have described naturalistic light exposure as a function of patient group, geographical location and season, and/or associated light exposure with health outcomes. Moreover, light dosimeters have been used to predict health-related outcome parameters in both clinical and non-clinical populations, rendering mixed findings (see, e.g., Ref. 52 ). The degree to which light dosimeters are useful for clinical contexts requires further research over long periods of time, using sensors that have been adequately assessed from verification and analytical validation perspectives. A key consideration is that the criteria used in characterizing a light dosimeter may differ between different types of trials and outcomes used.
Open questions in digital dosimetry
While the V3 framework provides a framework for considering the key technical and technological challenges for personalized sensor light dosimetry, there are additional questions that require theoretical and practical clarification.
To what extent is corneal light exposure a physiologically relevant quantity?
As stated earlier, retinal irradiance cannot be practically measured. Even when an observer is stationary and the scene constant, the light effectively reaching the eye and retina will be modulated in time by individual-level characteristics (such as head movements, eyelid closure, facial features, and pupil size18,85). Using some assumptions, e.g., a parametric model of pupil size as a function of corneal irradiance,86,87 the retinal irradiance could be reconstructed from corneal irradiance, but not in a spectrally selective way. In the limit, it is possible to conceive of two spatially very distinct scenarios yielding the same spatially integrated corneal irradiance: One very bright point light source in an otherwise dark field and one homogenous field. Importantly, the physiological effect of light varies with the region of the illuminated retina.88–93 At present, spatially resolved measurements are not generally available for wearable dosimetry to incorporate these effects.
Additionally, even if it is possible to estimate the spatially resolved and temporally defined pattern of α-opic radiances at the level of the retina, it is the photoreceptor signals that are relayed to targets in the brain that underlie the physiological effects of light. These may be subject to non-linearities, and may also be combined. For example, the ipRGCs receive input from the cones and rods,94,95 suggesting a pathway for integration of photoreceptor signals at the level of the retina already. For practical purposes, even if there is a biological capacity for the cones and rods to contribute to the physiological effects of light, there is converging evidence that melanopic quantities may be sufficient to predict them under most conditions.96–100
What is the usability and acceptability of digital dosimeters?
Different form factors for dosimeters place different demands on participants. Whether or not a portable sensor is usable will modify the extent to which it is fit-for-purpose in the field. While feasibility and acceptability concerns are typically considered during the planning phase of interventional clinical trials, it is critical to assess the usability, acceptability and equity of BioMeTs to be used in digital health and medicine. It is important to mention that there is typically a tradeoff between usability of light sensors and their accuracy in terms of approximation of corneal light exposure. For instance, sensors worn at the wrist might be preferred by users over the use of sensors mounted on a pair of glasses, but at the same time these sensors might render less accurate proxies of retinal light exposure as they capture light in a different plane and are more likely to be covered. Yet, usability is crucial for users’ adherence to the required (long-term) measurement protocol. Missing data, especially when systematic in specific contexts (such as when being in company of others and/or when spending time outdoors), might inhibit the collection of representative light exposure profiles in everyday-life settings.
Which metrics should be used?
An important consideration is that there is currently no standard set of metrics for summarizing time series data obtained with light sensors. This is particularly relevant for field assessments, with substantial inter- and intra-individual variations in light exposure profiles as a function of season, time of day and weather conditions and users’ location (e.g., indoors or outdoors) and behavior. 101 The lack of a standard set of metrics to quantify light exposure patterns challenges comparison across studies.
Moreover, it is not clear whether reported significant associations in the literature arose from a principled process, or whether parameters were tweaked until a significant association was found without statistical penalty. A potential solution is a multiverse approach, 102 preferably employed in a large data set, where a set of plausible choices of analytic parameters are systematically investigated and the robustness of a specific association is assessed. For studies reporting null findings, it could also be the case that the used aggregation missed important features of the light exposure profile that are particularly determinant in light-induced moderations in the outcome parameter.
How comparable are data from different groups and populations?
Portable sensor dosimeters can generate large amounts of data in different geographical locations or in groups with different demographic make-up. The development of common frameworks for storing and processing BioMeTs data will be a key in facilitating “big data” analyses of light dosimetry, and other types of data collected via digital sensors.103–105
How to handle non-wear periods and quality control?
The assessment of rest-activity cycles with actigraphy is often paired with the use of a sleep diary,106–111 sometimes indicating non-wear periods which can then be masked in later analysis steps. For light dosimetry, there are currently no standard instruments for capturing non-wear periods. In cases when the light-logging is integrated in a wrist-worn device also using actigraphy, the non-wear periods indicated in a sleep diary may be used for determining the periods on non-wear.
Quality control in general is an unsolved problem in light dosimetry, as there is no objective method for determining the congruency of corneal light exposure with measured light exposure. For example, sleeves may cover wrist-worn devices, and scarves may cover light dosimeters attached near the neck of a participant. A standard method of assessing the amount of invalid data needs to be developed.
Conclusion
Monitoring personalized light exposure through digital sensors is a key growing area in a variety of applications, ranging from its inclusion in associational studies, to light exposure being used as a medical intervention to treat, monitor or diagnose multiple conditions, or as a health-related outcome, moderator, or mediator in clinical trials. While there are now various light dosimeters available both for research use and commercially, it is not clear how to validate and verify them, and standard frameworks for the characterization of these light dosimeters do not necessarily apply. This article described the verification, analytic validation, and clinical validation (V3) framework to digital light-dosimetry sensors and highlighted ongoing challenges for portable digital dosimetry.
Footnotes
Acknowledgments
This publication is a result of collaborative research performed under the auspices of the Digital Medicine Society (DiMe). DiMe is a 510(c)(3) non-profit professional society for the digital medicine community, and is not considered a Sponsor of this work. All authors are members of DiMe who volunteered their contributions to this review. DiMe research activities are overseen by a Research Committee, the members of which were invited to comment on the manuscript prior to submission.
Contributorship
Conceptualization, Formal Analysis, Writing – original draft and Writing – review & editing: Manuel Spitschan, Karin Smolders, Benjamin Vandendriessche, Brinnae Bent, Jessie Bakker, Isaac R. Rodriguez-Chavez, Céline Vetter.
Data curation, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision and Validation: n/a.
Visualization: Manuel Spitschan.
Declaration of conflicting interests
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding
The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: During parts of this work, M.S. was supported by a Sir Henry Wellcome Postdoctoral Fellowship (Wellcome Trust, grant number 204686/Z/16/Z) and Linacre College, University of Oxford (Biomedical Sciences Junior Research Fellowship).
