Abstract
Background and aim:
Continuous glucose monitoring systems (CGMs) have been commercially available since 1999. However, automated insulin delivery systems may benefit from real-time inputs in addition to glucose. Continuous multi-analyte sensing platforms will meet this area of potential growth without increasing the burden of additional devices. We aimed to generate pilot data regarding the safety and function of a first-in-human, single-probe glucose/lactate multi-analyte continuous sensor.
Methods:
The investigational glucose/lactate continuous multi-analyte sensor (PercuSense Inc, Valencia, California) was inserted to the upper arms of 16 adults with diabetes, and data were available for analysis from 11 of these participants (seven female; mean [SD] = age 43 years [16]; body mass index [BMI] = 27 kg/m2 [5]). A commercially available Guardian 3 CGM (Medtronic, Northridge, California) was also inserted into the abdomen for comparison. All participants underwent a meal-test followed by an exercise challenge on day 1 and day 4 of wear. Performance was benchmarked against venous blood YSI glucose and lactate values.
Results:
The investigational glucose sensor had an overall mean absolute relative difference (MARD) of 14.5% (median = 11.2%) which improved on day 4 compared with day 1 (13.9% vs 15.2%). The Guardian 3 CGM had an overall MARD of 13.9% (median = 9.4%). The lactate sensor readings within 20/20% and 40/40% of YSI values were 59.7% and 83.1%, respectively.
Conclusions:
Our initial data support safety and functionality of a novel glucose/lactate continuous multi-analyte sensor. Further sensor refinement will improve run-in performance and accuracy.
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.