Abstract
Background:
This study evaluated differences in accuracy between the CONTOUR® NEXT EZ (EZ) blood glucose monitoring system (BGMS) and four other BGMSs [ACCU-CHEK® Aviva (ACAP), FreeStyle Freedom Lite® (FFL), ONE TOUCH® Ultra®2 (OTU2), and TRUEtrack® (TT)].
Methods:
Up to three capillary blood samples (
Results:
Across the tested glucose range, the EZ had the lowest MARD of 4.7%; the ACAP, FFL, OTU2, and TT had MARD values of 6.3%, 18.3%, 23.4%, and 26.2%, respectively. For samples with glucose concentrations <70 mg/dl, the EZ had the lowest MARD (0.65%), compared with the ACAP (2.5%), FFL (18.3%), OTU2 (22.4%), and TT (33.2%) systems.
Conclusions:
The EZ had the lowest MARD across the tested glucose ranges when compared with four other BGMSs when all samples were analyzed as well as when natural samples only were analyzed.