Abstract
Aims:
We developed a system for measuring glucose area under the curve (AUC) using minimally invasive interstitial fluid extraction technology (MIET). Sweat contamination during interstitial fluid glucose (IG) extraction affects the accuracy of glucose AUC measurement, because this technology uses extracted sodium ion levels as an internal standard. Therefore, we developed a sweat monitoring patch to reduce this effect and investigated its efficacy in volunteers undergoing oral glucose tolerance tests (OGTTs).
Results:
Good correlation between IG AUC measurements using MIET and reference AUCs measured using PG levels was confirmed over a wide AUC range (202–610 mg/h/dl) after correction for the sweat-induced error detected by the hydrogel patches on the nonpretreated skin. Strong correlation between IG AUC and peak glucose levels indicates that glucose spikes can be easily detected by this system.
Conclusion:
We confirmed the effectiveness of a sweat monitoring patch for precise AUC measurement using MIET. This novel, easy-to-use system has potential for glucose excursion evaluation in daily clinical practice.