Abstract
Background:
The purpose of this study was to design and evaluate a safety system for the artificial pancreas device system (APDS). Safe operation of the APDS is a critical task, where the safety system is engaged only as needed to ensure reliable operation without positive feedback to the controller.
Methods:
The Health Monitoring System (HMS) was designed as a modular system to ensure the safety of the APDS and the user. It was designed using a large set of ambulatory data and evaluated
Results:
The HMS was evaluated for clinical use with a 15 min prediction horizon. Retrospectively, 93.5% of episodes were detected with 2.9 false alarms per day. During
Conclusions:
The ability of the HMS to be an effective alert system that provides a safety layer to the APDS controller has been demonstrated in a clinical setting.