Abstract
Purpose
To report the unforeseen complication of total obstruction of a glaucoma drainage implant (GDI) tube lumen by white deposit material and to present a preliminary report identifying the composition of this material.
Methods
Two subjects with a high IOP due to total obstruction of a GDI tube were reviewed. Both patients had a long history with brinzolamide and timolol maleate eye drops. The GDI tube was swept with a 5-0 polypropylene suture stent in order to open the tube. The intraluminal solid sample was successfully collected from the implant tube in one patient. High-performance liquid chromatography-mass spectrometry (HPLC-MS) was used to determine the origin of the intraluminal sample.
Results
Intraluminal deposits containing components of antiglaucoma drugs e.g., timolol and brinzolamide are a rare cause of total obstruction of GDI tubes.
Conclusions
Our study describes a new cause of total
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.