Abstract
The aim of this work is to provide insights into multiple metrics clinical validation of deformable image registration and contour propagation methods in 4D lung radiotherapy planning. The following indices were analyzed and compared: Volume Difference (VD), Dice Similarity Coefficient (DSC), Positive Predictive Value (PPV) and Surface Distances (SD). The analysis was performed on three patient datasets, using as reference a ground-truth volume generated by means of Simultaneous Truth And Performance Level Estimation (STAPLE) algorithm from the outlines of five experts. Significant discrepancies in the quality assessment provided by the different metrics in all the examined cases were found. Metrics sensitivity was more evident in presence of image artifacts and particularly for tubular anatomical structures, such as esophagus or spinal cord. Volume Differences did not account for position and DSC exhibited criticalities due to its intrinsic symmetry (