BACKGROUND:
Previous research reporting factors influencing aerosol delivery in intubated patients has been largely focused on the endotracheal tube (ETT) during mechanical ventilation, with little comparative analysis of effect of types of artificial airways and their interfaces on aerosol delivery during spontaneous breathing. The purpose of this study was to compare aerosol delivery via tracheostomy tube (TT) and ETT, using interfaces such as T-piece, tracheostomy collar, and manual resuscitation bag.
METHODS:
A teaching manikin was intubated with either an ETT (8.0 mm inner diameter) and TT (8 mm inner diameter). Both bronchi were connected to a collecting filter, attached to a sinusoidal pump simulating the breathing pattern of a spontaneously breathing adult (tidal volume 450 mL, respiratory rate 20 breaths/min, inspiratory-expiratory ratio 1:2). Albuterol sulfate (2.5 mg/3 mL) was nebulized through a jet nebulizer, using each airway and interface as appropriate (
RESULTS:
A greater percentage of nominal dose was delivered via TT than ETT with both T-piece (13.79 ± 2.59% vs 9.05 ± 0.70%) and manual resuscitation bag (45.75 ± 1.8% vs 27.23 ± 8.98%,
CONCLUSION:
In this adult model of spontaneous ventilation, aerosol therapy through ETT was less efficient than TT, while the manual resuscitation bag was more efficient than T-piece or tracheostomy collar.