Aerosolized medications are routinely used for the treatment of critically ill patients. This paper reviews aerosol delivery devices with a focus on issues related to their performance in pulmonary critical care. Factors affecting aerosol drug delivery to mechanically ventilated adults and spontaneously breathing patients with artificial airways are reviewed. Device selection, optimum device technique, and unmet medical needs of aerosol medicine in pulmonary critical care are also discussed.
PhippsPR, GondaI. Droplets produced by medical nebulizers. Some factors affecting their size and solute concentration. Chest, 1990; 97(6):1327-1332.
9.
DolovichMB, DhandR. Aerosol drug delivery: developments in device design and clinical use. Lancet, 2011; 377(9770):1032-1045.
10.
RauJL. Design principles of liquid nebulization devices currently in use. Respir Care, 2002; 47(11):1257-1275; discussion 1275-1278.
11.
DhandR, MercierE. Effective inhaled drug administration to mechanically ventilated patients. Expert Opin Drug Deliv, 2007; 4(1):47-61.
12.
NikanderK, TurpeinenM, WollmerP. The conventional ultrasonic nebulizer proved inefficient in nebulizing a suspension. J Aerosol Med, 1999; 12(2):47-53.
13.
SteckelH, EskandarF. Factors affecting aerosol performance during nebulization with jet and ultrasonic nebulizers. Eur J Pharm Sci, 2003; 19(5):443-455.
14.
HarveyCJ, O'DohertyMJ, PageCJ, ThomasSH, NunanTO, TreacherDF. Comparison of jet and ultrasonic nebulizer pulmonary aerosol deposition during mechanical ventilation. Eur Respir J, 1997; 10(4):905-909.
15.
AriA. Nebulizers: an evaluation of nebulizers for better clinical practice. Eurasian J Pulmonol, 2014; 16:1-7.
16.
DhandR. New frontiers in aerosol delivery during mechanical ventilation. Respir Care, 2004; 49(6):666-677.
17.
DhandR. Nebulizers that use a vibrating mesh or plate with multiple apertures to generate aerosol. Respir Care, 47(12):1406-1416, 2002; discussion 1416-1408.
18.
AriA, AtalayOT, HarwoodR, SheardMM, AljamhanEA, FinkJB. Influence of nebulizer type, position, and bias flow on aerosol drug delivery in simulated pediatric and adult lung models during mechanical ventilation. Respir Care, 2010; 55(7):845-851.
19.
AriA, AreabiH, FinkJB. Evaluation of position of aerosol device in two different ventilator circuits during mechanical ventilation. Respir Care, 2010; 55(7):837-844.
20.
PitanceL, VecellioL, LealT, ReychlerG, ReychlerH, LiistroG. Delivery efficacy of a vibrating mesh nebulizer and a jet nebulizer under different configurations. J Aerosol Med Pulm Drug Deliv, 2010; 23(6):389-396.
21.
RubinBK, FinkJB. Optimizing aerosol delivery by pressurized metered-dose inhalers. Respir Care, 2005; 50(9):1191-1200.
22.
LabirisNR, DolovichMB. Pulmonary drug delivery. Part II: the role of inhalant delivery devices and drug formulations in therapeutic effectiveness of aerosolized medications. Br J Clin Pharmacol, 2003; 56(6):600-612.
23.
NewmanSP. Principles of metered-dose inhaler design. Respir Care, 2005; 50(9):1177-1190.
24.
RauJL. The inhalation of drugs: advantages and problems. Respir Care, 2005; 50(3):367-382.
25.
RubinBK. Air and soul: the science and application of aerosol therapy. Respir Care, 2010; 55(7):911-921.
26.
HessDR. Aerosol delivery devices in the treatment of asthma. Respir Care, 2008; 53(6):699-723; discussion 723-725.
27.
RauJL, DunlevyCL, HillRL. A comparison of inline MDI actuators for delivery of a beta agonist and a corticosteroid with a mechanically-ventilated lung model. Respir Care, 1998; 43:705-712.
28.
MacIntyreNR, SilverRM, MillerCW, SchulerF, ColemanRE. Aerosol delivery in intubated, mechanically ventilated patients. Crit Care Med, 1985; 13(2):81-84.
29.
FullerHD, DolovichMB, PosmituckG, PackWW, NewhouseMT. Pressurized aerosol versus jet aerosol delivery to mechanically ventilated patients. Comparison of dose to the lungs. Am Rev Respir Dis, 1990; 141(2):440-444.
30.
DhandR, TobinMJ. Inhaled bronchodilator therapy in mechanically ventilated patients. Am J Respir Crit Care Med, 1997; 156(1):3-10.
31.
FinkJ, DhandR, DuarteA, JenneJ, TobinM. Aerosol delivery from a metered-dose inhaler during mechanical ventilation. An in-vitro model. Am J Respir Crit Care Med, 1996; 154(2):382-387.
32.
FullerHD, DolovichMB, ChambersC, NewhouseMT. Aerosol delivery during mechanical ventilation: a predictive in-vitro lung model. J Aerosol Med, 1992; 5(4):251-259.
33.
FinkJB, DhandR, GrychowskiJ, FaheyPJ, TobinMJ. Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors. Am J Respir Crit Care Med, 1999; 159(1):63-68.
34.
DiotP, MorraL, SmaldoneGC. Albuterol delivery in a model of mechanical ventilation. Comparison of metered-dose inhaler and nebulizer efficiency. Am J Respir Crit Care Med, 1995; 152(4 Pt 1):1391-1394.
35.
RauJL, HarwoodRJ, GroffJL. Evaluation of a reservoir device for metered-dose bronchodilator delivery to intubated adults: an in-vitro study. Chest, 1992; 102(3):924-930.
36.
MorénF, AndersonJ. Fraction of dose exhaled after administration of pressurized inhalation aerosols. Int J Pharm, 1980; 6(3-4):295-300.
37.
GoodeML, FinkJB, DhandR, TobinMJ. Improvement in aerosol delivery with helium-oxygen mixtures during mechanical ventilation. Am J Respir Crit Care Med, 2001; 163(1):109-114.
38.
O'DohertyMJ, ThomasSH, PageCJ, TreacherDF, NunanTO. Delivery of a nebulized aerosol to a lung model during mechanical ventilation. Effect of ventilator settings and nebulizer type, position, and volume of fill. Am Rev Respir Dis, 1992; 146(2):383-388.
39.
ThomasSH, O'DohertyMJ, PageCJ, TreacherDF, NunanTO. Delivery of ultrasonic nebulized aerosols to a lung model during mechanical ventilation. Am Rev Resp Dis, 1993; 148(4):872-877.
40.
ThomasSH, O'DohertyMJ, FidlerHM, PageCJ, TreacherDF, NunanTO. Pulmonary deposition of a nebulised aerosol during mechanical ventilation. Thorax, 1993; 48(2):154-159.
41.
HessDR, FisherD, WilliamsP, PoolerS, KacmarekRM. Medication nebulizer performance. Effects of diluent volume, nebulizer flow, and nebulizer brand. Chest, 1996; 110(2):498-505.
42.
HarveyCJ, O'DohertyMJ, PageCJ, ThomasSH, NunanTO, TreacherDF. Effect of a spacer on pulmonary aerosol deposition from a jet nebulizer during mechanical ventilation. Thorax, 1995; 50(1):50-53.
43.
NewmanSP. How well do in vitro particle size measurements predict drug delivery in vivo?. J Aerosol Med, 1998; 11(Suppl 1):S97-104.
44.
DhandR, SohalH. Pulmonary drug delivery system for inhalation therapy in mechanically ventilated patients. Expert Rev Med Devices, 2008; 5(1):9-18.
45.
WaldrepJC, BerlinskiA, DhandR. Comparative analysis of methods to measure aerosols generated by a vibrating mesh nebulizer. J Aerosol Med, 2007; 20(3):310-319.
46.
Van DykeRE, NikanderK. Delivery of iloprost inhalation solution with the HaloLite, Prodose, and I-neb adaptive aerosol delivery systems: an in vitro study. Respir Care, 2007; 52(2):184-190.
47.
PolliGP, GrimWM, BacherFA, YunkerMH. Influence of formulation on aerosol particle size. J Pharm Sci, 1969; 58(4):484-486.
48.
KimCS, TrujilloD, SacknerMA. Size aspects of metered-dose inhaler aerosols. Am Rev Respir Dis, 1985; 132(1):137-142.
49.
EhrmannS, Roche-CampoF, Sferrazza PapaGF, IsabeyD, BrochardL, Apiou-SbirleaG. Aerosol therapy during mechanical ventilation: an international survey. Intensive Care Med, 2013; 39(6):1048-1056.
50.
MacNeishCF, MeisnerD, ThibertR, KelemenS, VadasEB, CoatesAL. A comparison of pulmonary availability between Ventolin (albuterol) nebules and Ventolin (albuterol) respirator solution. Chest, 1997; 111(1):204-208.
51.
BerlinskiA, WaldrepJC. Nebulized drug admixtures: effect on aerosol characteristics and albuterol output. J Aerosol Med, 2006; 19(4):484-490.
52.
CoatesAL, MacNeishCF, MeisnerD, KelemenS, ThibertR, MacDonaldJ, VadasE. The choice of jet nebulizer, nebulizing flow, and addition of albuterol affects the output of tobramycin aerosols. Chest, 1997; 111(5):1206-1212.
53.
FlamentMP, LetermeP, BurnoufT, GayotA. Influence of formulation on jet nebulisation quality of α1 protease inhibitor. Int J Pharm, 1999; 178(1):101-109.
54.
LeeTY, ChenCM, LeeCN, ChiangYC, ChenHY. Compatibility and osmolality of inhaled N-acetylcysteine nebulizing solution with fenoterol and ipratropium. Am J Health Syst Pharm, 2005; 62(8):828-833.
55.
JosephJC. Compatibility of nebulizer solution admixtures. Ann Pharmacother, 1997; 31(4):487-489.
56.
KaminW, SchwabeA, KrämerI. Inhalation solutions: which ones are allowed to be mixed? Physico-chemical compatibility of drug solutions in nebulizers. J Cyst Fibros, 2006; 5(4):205-213.
57.
McKenzieJE, Cruz-RiveraM. Compatibility of budesonide inhalation suspension with four nebulizing solutions. Ann Pharmacother, 2004; 38(6):967-972.
58.
AlvineGF, RodgersP, FitzsimmonsKM, AhrensRC. Disposable jet nebulizers. How reliable are they?. Chest, 1992; 101(2):316-319.
59.
LoffertDT, IkleD, NelsonHS. A comparison of commercial jet nebulizers. Chest, 1994; 106(6):1788-1792.
60.
TerzanoC, PetroianniA, ParolaD, RicciA. Compressor/nebulizer differences in the nebulization of corticosteroids. The CODE study (Corticosteroids and Devices Efficiency). Eur Rev Med Pharmacol Sci, 2007; 11(4):225-237.
61.
WaldrepJC, KeyhaniK, BlackM, KnightV. Operating characteristics of 18 different continuous-flow jet nebulizers with beclomethasone dipropionate liposome aerosol. Chest, 1994; 105(1):106-110.
62.
CravenDE, GoularteTA, MakeBJ. Contaminated condensate in mechanical ventilator circuits: a risk factor for nosocomial pneumonia. Am Rev Respir Dis, 1984; 129(4):625-628.
PhillipsGD, MillardFJ. The therapeutic use of ultrasonic nebulizers in acute asthma. Respir Med, 1994; 88(5):387-389.
65.
DhandR, GunturVP. How best to deliver aerosol medications to mechanically ventilated patients. Clin Chest Med, 2008; 29(2):277-296.
66.
DhandR. Aerosol delivery during mechanical ventilation: from basic techniques to new devices. J Aerosol Med Pulm Drug Deliv, 2008; 21(1):45-60.
67.
WaldrepJC, DhandR. Advanced nebulizer designs employing vibrating mesh/aperture plate technologies for aerosol generation. Curr Drug Deliv, 2008; 5(2):114-119.
68.
HurleyPK, SmyeSW, CunliffeH. Assessment of antibiotic aerosol generation using commercial jet nebulizers. J Aerosol Med, 1994; 7(3):217-228.
69.
DhandR. Aerosol therapy during mechanical ventilation: getting ready for prime time. Am J Respir Crit Care Med, 2003; 168(10):1148-1149.
70.
MillerDD, AminMM, PalmerLB, ShahAR, SmaldoneGC. Aerosol delivery and modern mechanical ventilation: in vitro/in vivo evaluation. Am J Respir Crit Care Med, 2003; 168(10):1205-1209.
71.
HughesJ, SaezJ. Effects of nebulizer mode and position in a mechanical ventilator circuit on dose efficiency. Respir Care, 1987; 32(12):1131-1135.
72.
Sidler-MoixAL, DolciU, Berger-GryllakiM, PannatierA, CottingJ, Di PaoloER. Albuterol delivery in an in vitro pediatric ventilator lung model: comparison of jet, ultrasonic and mesh nebulizers. Pediatr Crit Care Med, 2013; 14(2):e98-e102.
73.
WanGH, LinHL, FinkJB, ChenYH, WangWJ, ChiuYC, et al. In vitro evaluation of aerosol delivery by different nebulization modes in pediatric and mechanical ventilator modes. Respir Care, 2014; 59(10):1494-1500.
74.
BerlinskiA, WillisJR. Albuterol delivery by 4 different nebulizers placed in 4 different positions in a pediatric ventilator in vitro model. Respir Care, 2013; 58(7):1124-1133.
75.
MoraineJJ, TruflandierK, VandenbergenN, BerréJ, MélotC, VincentJL. Placement of the nebulizer before the humidifier during mechanical ventilation: Effect on aerosol delivery. Heart Lung, 2009; 38(5):435-439.
76.
FinkJ. Metered-dose inhalers, dry powder inhalers and transitions. Respir Care, 2000; 45(6):623-635.
77.
EverardML, DevadasonSG, SummersQA, Le SouëfPN. Factors affecting total and “respirable” dose delivered by a salbutamol metered dose inhaler. Thorax, 1995; 50(7):746-749.
78.
O'CallaghanC, LynchJ, CantM, RobertsonC. Improvement in sodium cromoglycate delivery from a spacer device by use of an antistatic lining, immediate inhalation, and avoiding multiple actuations of drug. Thorax, 1993; 48(6):603-606.
79.
BarryPW, O'CallaghanC. The effect of delay, multiple actuations and spacer static charge on the in vitro delivery of budesonide from the Nebuhaler. Br J Clin Pharmacol, 1995; 40(1):76-78.
80.
ClarkDJ, LipworthBJ. Effect of multiple actuations, delayed inhalation and antistatic treatment on the lung bioavailability of salbutamol via a spacer device. Thorax, 1996; 51(10):981-984.
81.
RauJL, ZhuY. Reservoir design and dose availability with long-term metered dose inhaler corticosteroid use. J Aerosol Med, 1998; 11(1):15-26.
82.
BishopMJ, LarsonRP, BuschmanDL. Metered dose inhaler aerosol characteristics are affected by the endotracheal tube actuator/adapter used. Anesthesiology, 1990; 73(6):1263-1265.
83.
WilliamsRO3rd, PatelAM, BarronMK, RogersTL. Investigation of some commercially available spacer devices for the delivery of glucocorticoid steroids from a pMDI. Drug Dev Ind Pharm, 2001; 27(5):401-412.
84.
FullerHD, DolovichMB, TurpieFH, NewhouseMT. Efficiency of bronchodilator aerosol delivery to the lungs from the metered dose inhaler in mechanically ventilated patients. A study comparing four different actuator devices. Chest, 1994; 105(1):214-218.
85.
DhandR, DuarteAG, JubranA, JenneJW, FinkJB, FaheyPJ, TobinMJ. Dose response to bronchodilator delivered by metered-dose inhaler in ventilator supported patients. Am J Respir Crit Care Med, 1996; 154(2):388-393.
86.
DhandR, TobinMJ. Bronchodilator delivery with metered-dose inhalers in mechanically-ventilated patients. Eur Respir J, 1996; 9(3):585-595.
87.
O'RiordanTG, PalmerLB, SmaldoneGC. Aerosol deposition in mechanically ventilated patients. Optimizing nebulizer delivery. Am J Respir Crit Care Med, 1994; 149(1):214-219.
88.
GuerinC, FassierT, BayleF, LemassonS, RichardJC. Inhaled bronchodilator administration during mechanical ventilation: how to optimize it and for which clinical benefit?. J Aerosol Med Pulm Drug Deliv, 2008; 21(1):85-96.
89.
HessDR, DillmanC, KacmarekRM. In vitro evaluation of aerosol bronchodilator delivery during mechanical ventilation: pressure-control vs. volume control ventilation. Intensive Care Med, 2003; 29(7):1145-1150.
90.
AndersenJB, KlausenNO. A new mode of administration of nebulized bronchodilator in severe bronchospasm. Eur J Respir Dis Suppl, 1982; 119:97-100.
91.
FinkJB, TobinMJ, DhandR. Bronchodilator therapy in mechanically ventilated patients. Respir Care, 1999; 44(1):53-69.
92.
CroganSJ, BishopMJ. Delivery efficiency of metered dose aerosols given via endotracheal tubes. Anesthesiology, 1989; 70(6):1008-1010.
93.
GarnerSS, WiestDB, BradleyJW. Albuterol delivery by metered-dose inhaler with a pediatric mechanical ventilatory circuit model. Pharmacotherapy, 1994; 14(2):210-214.
94.
RyanSN, RankinN, MeyerE, WilliamsR. Energy balance in the intubated human airway is an indicator of optimal gas conditioning. Crit Care Med, 2002; 30(2):355-361.
95.
SchulzeA. Respiratory gas conditioning in infants with an artificial airway. Semin Neonatol, 2002; 7(5):369-377.
96.
LangeCF, FinlayWH. Overcoming the adverse effect of humidity in aerosol delivery via pressurized metered-dose inhalers during mechanical ventilation. Am J Respir Crit Care Med, 2000; 161(5):1614-1618.
97.
LinHL, FinkJB, ZhouY, ChengYS. Influence of moisture accumulation in inline spacer on delivery of aerosol using metered-dose inhaler during mechanical ventilation. Respir Care, 2009; 54(10):1336-1341.
98.
BowersW. Effects of heat and moisture exchangers designed to allow aerosol delivery on aerosol deposition in simulated mechanically ventilated adults. Atlanta, GA: Georgia State University; 2010.
99.
SvartengrenM, AndersonM, PhilipsonK, CamnerP. Human lung deposition of particles suspended in air or in helium/oxygen mixture. Exp Lung Res, 1989; 15(4):575-585.
100.
AriA, FinkJ. Aerosol drug delivery administration with helium-oxygen (heliox) mixtures: an overview. Curr Respir Med Rev, 2010; 6(2):80-85.
101.
DuarteAG. Inhaled bronchodilator administration during mechanical ventilation. Respir Care, 2004; 49(6):623-634.
102.
TassauxD, JollietP, ThouretJM, RoeselerJ, DorneR, ChevroletJC. Calibration of seven ICU ventilators for mechanical ventilation with helium-oxygen mixtures. Am J Respir Crit Care Med, 1999; 160(1):22-32.
103.
TakayaT, TakeyamaK, TakiguchiM. The efficiency of β2-agonist delivery through tracheal tubes with the metered-dose inhaler: an in vitro study. J Anesth, 2002; 16(4):284-288.
104.
AhrensRC, RiesRA, PopendorfW, WieseJA. The delivery of therapeutic aerosols through endotracheal tubes. Pediatr Pulmonol, 1986; 2(1):19-26.
105.
O'RiordanTG, GrecoMJ, PerryRJ, SmaldoneGC. Nebulizer function during mechanical ventilation. Am Rev Respir Dis, 1992; 145(5):1117-1122.
106.
PiccuitoCM, HessDR. Albuterol delivery via tracheostomy tube. Respir Care, 2005; 50(8):1071-1076.
107.
AriA, HarwoodRJ, SheardMM, FinkJB. An in vitro evaluation of aerosol delivery through tracheostomy and endotracheal tubes using different interfaces. Respir Care, 2012; 57(7):1066-1070.
108.
BerlinskiA. Nebulized albuterol delivery in a model of spontaneously breathing children with tracheostomy. Respir Care, 2013; 58(12):2076-2086.
109.
BerlinskiA, ChavezA. Albuterol delivery via metered dose inhaler in a spontaneously breathing pediatric tracheostomy model. Pediatr Pulmonol, 2013; 48(10):1026-1034.
110.
MirzaS, HopkinsonL, MalikTH, WillattDJ. The use of inhalers in patients with tracheal stomas or tracheostomy tubes. J Laryngol Otol, 1999; 113(8):762-764.
111.
NakhlaV. A homemade modification of a spacer device for delivery of bronchodilator or steroid therapy in patients with tracheostomies. J Laryngol Otol, 1997; 111(4):363-365.
112.
ParkesSN, BerstenAD. Aerosol kinetics and bronchodilator efficacy during continuous positive airway pressure delivered by face mask. Thorax, 1997; 52(2):171-175.
113.
MacIntyreN. Aerosol delivery through an artificial airway. Respir Care, 2002; 47(11):1279-1285; discussion 1285-1289.
114.
DhandR. Special problems in aerosol delivery: artificial airways. Respir Care, 2000; 45(6):636-645.
115.
PitanceL, VecellioL, DelvalG, ReychlerG, ReychlerH, LiistroG. Aerosol delivery through tracheostomy tubes: an in vitro study. J Aerosol Med Pulm Drug Deliv, 2013; 26(2):76-83.
116.
DolovichMB, AhrensRC, HessDR, AndersonP, DhandR, RauJL, et al. Device selection and outcomes of aerosol therapy: evidence-based guidelines: American College of Chest Physicians/American College of Asthma, Allergy, and Immunology. Chest, 2005; 127(1):335-371.
117.
LaubeBL, JanssensHM, de JonghFH, DevadasonSG, DhandR, DiotP, et al. What the pulmonary specialist should know about the new inhalation therapies. Eur Respir J, 2011; 37(6):1308-1331.
118.
MarikP, HoganJ, KrikorianJ. A comparison of bronchodilator therapy delivered by nebulization and metered-dose inhaler in mechanically ventilated patients. Chest, 1999; 115(6):1653-1657.
119.
ElyEW, BakerAM, EvansGW, HaponikEF. The distribution of costs of care in mechanically ventilated patients with chronic obstructive pulmonary disease. Crit Care Med, 2000; 28(2):408-413.
120.
DhandR. Inhalation therapy in invasive and noninvasive mechanical ventilation. Curr Opin Crit Care, 2007; 13(1):27-38.
121.
DhandR. Inhalation therapy with metered-dose inhalers and dry powder inhalers in mechanically ventilated patients. Respir Care, 2005; 50(10):1331-1334; discussion 1344-1345.
122.
DuarteAG, MomiiK, BidaniA. Bronchodilator therapy with metered dose inhaler and spacer versus nebulizer in mechanically ventilated patients: comparison of magnitude and duration of response. Respir Care, 2000; 45(7):817-823.
123.
ChangLH, HonidenS, HaithcockJA, DasAM, ShortKA, NiermanDM, CarsonSS. Utilization of bronchodilators in ventilated patients without obstructive airways disease. Respir Care, 2007; 52(2):154-158.
124.
CravenDE, LichtenbergDA, GoularteTA, MakeBJ, McCabeWR. Contaminated medication nebulizers in mechanical ventilator circuits. Source of bacterial aerosols. Am J Med, 1984; 77(5):834-838.
125.
HamillRJ, HoustonED, GeorghiouPR, WrightCE, KozaMA, CadleRM, et al. An outbreak of Burkholderia cepacia respiratory tract colonization and infection associated with nebulized albuterol therapy. Ann Intern Med, 1995; 122(10):762-766.
126.
HamillRJ, HoustonED, GeorghiouPR, WrightCE, KozaMA, CadleRM, et al. Respiratory tract colonization and infection associated with nebulized albuterol therapy. Ann Intern Med, 1995; 122(10):762-766.
127.
FinkJB, DhandR, DuarteAG, JenneJW, TobinMJ. Aerosol delivery from a metered-dose inhaler during mechanical ventilation: an in-vitro model. Am J Respir Crit Care Med, 1996; 154(2 Pt 1):382–387.
128.
MouloudiE, KatsanoulasK, AnastasakiM, HoingS, GeorgopoulosD. Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of tidal volume. Intensive Care Med, 1999; 25(11):1215–1221.
129.
AriA, AreabiH, FinkJB. Evaluation of aerosol generators at 3 locations in humidified and nonhumidified circuits during adult mechanical ventilation. Respir Care, 2010; 55(7):837–844.
130.
HessDR, DillmanC, KacmarekRM. In vitro evaluation of aerosol bronchodilator delivery during mechanical ventilation: pressure-control vs volume-control ventilation. Intensive Care Med, 2003; 29(7):1145–1150.
131.
HessDR, FisherD, WilliamsP, PoolerS, KacmarekRM. Medication nebulizer performance. Effects of diluent volume, nebulizer flow, and nebulizer brand. Chest, 1996; 110(2):498–505.
132.
BerlinskiA, WillisJR. Albuterol delivery by 4 different nebulizers placed in 4 different positions in a pediatric ventilator in vitro model. Respir Care, 2013; 58(7):1124–1133.
National Heart, Lung, and Blood Institute Acute Respiratory Distress Syndrome (ARDS) Clinical Trials Network, MatthayMA, BrowerRG, CarsonS, DouglasIS, EisnerM, et al. Randomized, placebo-controlled clinical trial of an aerosolized β2-agonist for treatment of acute lung injury. Am J Respir Crit Care Med, 2011; 184(5):561–568.
135.
AlzahraniW, HarwoodR, FinkJB, GoodfellowL, AriA. Comparison of albuterol delivery during high frequency oscillatory ventilation and conventional mechanical ventilation of a simulated adult. Respir Care, 2010; 55(11):1576.
136.
DuarteAG. Inhaled bronchodilator administration during mechanical ventilation. Respir Care, 2004; 49(6):623–634.