Abstract
Background
To better understand the spectrum of overfill reports and their corresponding clinical severity and etiology, we conducted a review of overfill reports from the Manufacturer and User Facility Device Experience (MAUDE) database, which is within the Food and Drug Administration (FDA) Web site (www.fda.gov).
Method
We searched the MAUDE database for events related to overfill reports between 1 January 1995 and 31 December 2008 and recorded drain volume (DV)/fill volume (FV), or DV/FV, and clinical symptoms and signs associated with the overfill report.
Results
Among 462 MAUDE reports with a possible overfill event, 440 reports (95.2%) with a confirmed overfill event contained sufficient information to ascertain the clinical severity of the event. The number of reports with a clinical severity rating of minor, moderate, major, or death was 331, 71, 28, and 10, respectively. The median (range) DV/FV for a subgroup of 292 reports with a clinical severity rating of minor, moderate, major, or death was 1.63 (1.06 - 4.29), 1.71 (1.08 - 5.87), 2.14(1.64 - 2.61), and 2.50 (2.28 - 3.33), respectively. Insufficient drain accounted for a majority of overfill reports.
Conclusion
Our analysis of reports from the MAUDE database suggests an association between DV/FV and clinical severity of the reported overfill event, as well as significant patient-to-patient variability with respect to intra peritoneal volume tolerance.