Abstract
BACKGROUND:
It has not yet been tested whether averaged gain values and the presence of pathological saccades are significantly altered by manual data selection or if data selection only done by the incorporated software detection algorithms provides a reliable data set following v-HIT testing.
OBJECTIVE:
The primary endpoint was to evaluate whether the averaged gain values of all six SCCs are significantly altered by manual data selection with two different v-HIT systems.
METHOD:
120 subjects with previously neither vestibular nor neurological disorders underwent four separate tests of all six SCCs with either EyeSeeCam® or ICS Impulse®. All v-HIT test reports underwent manual data selection by an experienced ENT Specialist with deletion of any noise and/or artifacts. Generalized estimating equations were used to compare averaged gain values based on unsorted data with averaged gain values based on the sorted data.
RESULTS:
EyeSeeCam®: Horizontal SCCs: The estimate and the
CONCLUSION:
None of the two v-HIT systems revealed any clinically important effects of manual data selection. However, 21 fewer tests were considered pathological after manual data selection.