Abstract
BACKGROUND:
A multi-pod catheter (MPC) is a large drainage catheter that can house multiple smaller retractable (MPC-R) and deployable catheters (MPC-D) within the body.
OBJECTIVE:
The drainage capabilities and resistance to clogging of a novel MPC have been assessed.
METHODS:
The drainage capabilities are evaluated by placing the MPC in a bag of either a non-clogging (H
RESULTS:
In the non-clogging medium, MPC-D had a slightly higher MaxDV than MPC-R, and higher flow rate than CTC and MPC-R. Moreover, MPC-D needed less TTD200 than MPC-R. In the clogging medium, MPC-D had a higher MaxDV than CTC and OTC, higher flow rate, and faster TTD200 than CTC. However, comparison with MPC-R showed no significant difference.
CONCLUSION:
The novel catheter may offer superior drainage compared to the single-lumen catheter in a clogging medium, implying various clinical applications, particularly when clogging is a potential risk. Further testing may be required to simulate various clinical scenarios.
Get full access to this article