Abstract
INTRODUCTION:
Dose evaluation with two Monte Carlo codes using patient specific voxel phantom is presented in this paper. We employ both MCNPX and DOSXYZnrc to perform dosimetry for mathematical voxel phantoms generated by our in-house developed voxel phantom generator and EGSnrc/CTCreate respectively.
MATERIAL AND METHOD:
Our case study was a 2.5 × 2.4 × 2.4 cm3 tumor in the middle lobe of right lung of a male patient exposed to 6MV parallel beam. In order to compare these Monte Carlo codes with together gross tumor volume (GTV) and organ at risks (OAR) doses and dose volume histograms (DVH) were calculated.
RESULTS:
Comparing the mean absorbed dose results (in Gy) from both codes indicates that gross tumor volume, heart and spinal cord have 2% to 10% difference. The 10% difference between the codes were from the spinal cord region where was not in the therapy beam and it just received the scatter radiation. The dose volume DVH obtained from DOSXYZnrc results demonstrate a milder slope compared with MCNPX DVHs.
CONCLUSION:
It was revealed that MCNPX has some advantages in comparison to DOSXYZnrc, but it is important to consider that for equal precision in voxel dosimetry calculation, DOSXYZnrc runs faster than MCNPX and it is a great advantage.