Abstract
BACKGROUND:
Fatigability measured and derived parameters are commonly used in research and clinical contexts to characterize performance during isokinetic fatigue protocols. The fatigability measured parameters are the best repetition, the total and the partial sums while the fatigability derived parameters are ratios and formula-based parameters.
OBJECTIVES:
To comprehensively evaluate the reproducibility of fatigability measured and derived parameter and to determine which of these is/are sufficiently interpretable for assessment of knee muscles.
METHODS:
Eighteen sedentary men underwent three isokinetic knee fatigability testing sessions with 7–10 days of rest between each session. Peak moment (PM) and maximal work (MW) were computed for each repetition and analyzed to calculate 54 measured and derived parameters. Relative (Intra-class Correlation Coefficients – ICC) and absolute (Coefficient of Variation of Method Error – CV
RESULTS:
All fatigability measured parameters (save one) were associated with a high relative and absolute reproducibility for knee extensors (ICC
CONCLUSIONS:
All fatigability measured parameters may be used for assessing knee extensors fatigue with either PM or MW; for assessing knee flexors, no measured parameters can be utilized. Above all, knee fatigability derived parameters, either PT- or MW-based, should not be used, for both the extensors and the flexors of the knee, due to clinically unacceptable reproducibility.
Keywords
Introduction
Based on its reproducibility and sensitivity [1, 2] in research and clinical contexts [3], isokinetic testing has become a gold standard in muscular strength evaluations. Commonly, isokinetic devices are used to assess strength at various velocities by means of short duration testing. However, in some cases, the evaluation of muscle endurance to fatigue may be more specific than that of strength; for instance, in athletes who mainly use lactic anaerobic metabolism to perform or in patients with neuromuscular pathology [4, 5].
As defined by Bigland-Ritchie et al., fatigue refers to “any reduction in the capacity of the total neuromuscular system to generate force, regardless the force level deployed” [6, 7]. Generally, the knee fatigability protocol consists of a defined number of reciprocal maximal concentric contractions or in the largest number of repetitions in a given period of time, all performed at a prescribed velocity, e.g. 180
In the clinical context practitioners commonly use some fatigability derived parameters which, in some cases, are provided by the dynamometer’s dedicated software. For instance, the Biodex software compares the first and the last third of repetitions in the set to obtain a percentage decline. The fatigability measured parameters are the best performance, the total and partials sums while the fatigability derived parameters are the quotient between the
However, the question is still open on which isokinetic fatigability parameter(s) may be recommended for use in evaluating the knee muscles. The first step in this determination is to clarify their intra-dynamometer reproducibility. Although several studies had assessed the reproducibility of knee isokinetic fatigability protocols [6, 9, 12, 13, 14, 32, 33, 34, 35, 36, 37, 38, 39], to our knowledge, none has done so in terms of both the relative and the absolute reproducibility of the fatigability parameters. Therefore, the goal of our study was to evaluate, as comprehensively as possible, the intra-dynamometer reproducibility of fatigability measured and derived parameters so as to determine which have sufficient reproducibility for application in clinical and research environments.
Methods
Subjects
Eighteen recreationally men, with no history of (major) knee injury, practicing less than two hours of physical activity per week and not involved in specific muscle training of the lower limbs were included in this study. Any drug or medication intake was an exclusion criterion. This study, approved by the Ethics Committee of the University and Faculty Hospital, was conducted in conformity with ethical standards and (inter) national laws [40] and with the Helsinki declaration of 1975, as revised in 2000. All subjects were informed about the protocol, submitted to a medical examination and signed an information and consent form. Age, size and body weight means
Experimental design
Three isokinetic testing sessions were carried out, with seven to ten days of rest between each session, for each subject under the supervision of a single physical therapist. To avoid possible effects of a residual fatigue due to recent exercises, subjects were asked to refrain from physical activity for forty-eight hours before each evaluation session. In order to standardize the testing session conditions, subjects were asked:
to follow a stable sleep schedule in the three days before the sessions; not to eat any meal in the three hours preceding the sessions; not to drink coffee, tea or any other energy drink the day of the sessions.
The evaluations were performed on a Biodex dynamometer (Biodex Medical Systems 3, Shirley, NY). This dynamometer has an excellent reliability with an ICC of 0.99 for position, peak moment and velocity [41]. We tested only the dominant leg, defined as kicking leg. The test position was in sitting, with the backrest raised at 85
Prior to testing, subjects were instructed to push up and pull down until the dynamometer stopped, in order to achieve a full RoM. To ensure familiarization and complete the set-up, the subjects were making submaximal but progressively intensified concentric contractions intensity at 120
In this study we used the same database as did Bosquet et al. [6] who studied the effects of lengthening the isokinetic evaluation of muscle endurance to fatigue. The subjects underwent the formal test, which consisted of 50 consecutive maximal reciprocal flexion-extension concentric contractions at 180
Data analysis
Peak moment (PM in N
Knee parameters
Knee parameters
Note:
Relative reproducibility was determined using the ICC
Absolute reproducibility was assessed using the Standard Error of the Mean (SEM), considered as expected trial-to-trial noise in the data [50]. The Minimum Detectable Change (MDC), based on SEM, which is defined as minimal amount of change outside of error that reflects true change by a subject between two evaluations, was also calculated. SEM and MDC are expressed in the unit (N
ICC (Eq. (2.4)), SEM (Eq. (2)) and MDC at the 95% confidence level (Eq. (3)) were computed from the table of a two-way analysis of variance (trials
where
Where
The absolute (Eq. (4)) and relative (Eq. (5)) variability are calculated using the following equations:
where
The Coefficient of Variation of Method Error (CV
Where SEM is defined previously (Eq. (2)) and where mean of parameter is the mean value of the measured or derived parameter’s value (from the three tests occasions).
The linear correlation between two variables was measured by the Pearson’s r product-moment correlation coefficient (PCC) and linear regression analyses with coefficients of determination associated (r
Statistical significance was set at the level of p-value
The PM and MW mean values and standard deviation from the three testing sessions for the knee extensors and flexors are presented in Table 2.
Knee mean value
standard deviation for testing sessions 1–3
Knee mean value
Absolute and relative variability mean value
Knee extensors parameters Peak moment & Maximal Work mean value
Note:
The reproducibility of data collected on knee extensors is presented in Table 4.
The fatigability measured parameters, with the exception of partial sum #26–#30 (
The relative reproducibility of the fatigability derived parameters was (at best) moderate: the ICC, for PM and MW, was
The r
Relative and absolute variability between consecutive repetitions showed low reproducibility, with ICC less than 0.4 for knee extensors, MDC score higher than 70.0% and none of the SEM scores met the criterion for acceptable precision, for PM as well as MW (see Table 3). The relative variability was nevertheless strongly and positively correlated with the progress of the testing session (PCC
Reproducibility of fatigability parameters for knee flexors
The data reproducibility for knee flexors is presented in Table 5.
Knee flexors parameters peak moment and maximal work mean value
standard deviation, ICC, ICC CI 95%, CV, SEM (expressed in N
m for PM and in J for MW) and MDC
Knee flexors parameters peak moment and maximal work mean value
Note:
Only the total sum (
All fatigability derived parameters had an ICC lower than 0.52, whether for PM or MW, and, therefore, had a weak (to moderate) relative reproducibility. The MDC of all derived parameters were higher than 55.0% (for the PM and the MW) except for the quotient of the total sum by the best repetition (
The reproducibility of the fatigability measured and derived parameters for the flexors was lower than for the extensors either for the PM or the MW (the ICC was steadily equal or higher while the CV
The r
The PCC between the extensors and the flexors was 0.807 and 0.796 respectively for the PM and MW (with
Relative and absolute variability between consecutive repetitions showed low reproducibility with ICC less than 0.5 for knee flexors, MDC score higher than 60.0% and none of the SEM scores met the criterion for acceptable precision, for PM and MW (see Table 3). The relative variability was nevertheless strongly and positively correlated with the progress of the testing session (PCC
Although Lambert et al. [61], Pincivero et al. [13, 14] and Saenz et al. [33] demonstrated that the measured parameters of fatigability assessment are reliable by contrast with the derived parameters, we have to specify that the different authors have analyzed only one derived parameter by research (respectively, ratio between the 15 first and the 15 last repetitions, between the 5 first and the 5 last repetitions, between the 10 first and the 10 last repetitions and between the 13 first and the 13 last repetitions). Respectively in 2000 and 2011, Pincivero et al. [13] and Ferriero et al. [62] have briefly questioned the utilization of the fatigability derived parameters for the assessment of the decrease in performance but, to our knowledge, no previous research has comprehensively studied the intra-dynamometer reproducibility of those fatigability derived parameters employed for the isokinetic evaluation of knee fatigability in concentric conditions. In the present investigation we conducted a reproducibility study that included the most exhaustive number of logical combinations of measured and derived parameters regularly provided by the isokinetic devices or used by clinicians or researchers, as listed in the data analysis section.
For the measured parameters, our data corroborate previous research: almost all showed excellent relative reproducibility for knee extensors and flexors, whether for PM or MW. However: (1) the measured parameters (best repetitions, total and partial sums) have a sufficient absolute and relative reproducibility (whether for PM or MW) for evaluation also only for the extensors even if a careful consideration must be given to the partial sums at the end of testing session (#21–#30, #21–#25 and #26–#30). The lower reproducibility of the partial sums at the end of testing session can be partly explained by the patient motivation and/or the arduousness of the fatigability evaluation. This will be discussed in more detail below along with the relative and absolute variability; (2) despite the fact that their relative reproducibility is high, due to their absolute reproducibility which did not meet the criterion for acceptable precision, no fatigability measured parameters can be recommended for knee flexors.
As explained by Kean et al. [63], the MDC is “particularly useful when determining whether an observed change in score between test sessions is truly a change in subject status or is merely error (noise) in the measurement” [50]. Our SEM and MDC results for the measured parameters were, despite the appearance to the contrary, lower than a previous research that assessed the reliability of the same isokinetic device employed in the present study (Biodex System 3): observed SEM scores of about 11.8 N
The lower relative and absolute reproducibility of the knee flexors, demonstrated by the higher scores of ICC or ANOVA main effect p-value and the lower scores of CV
As mentioned by Bosquet et al. [6], PM and MW are not affected identically by muscle fatigue induction. Indeed, the PM and the MW are very strongly correlated to one another (PCC around 0.98), whether it is for the extensors or the flexors, but the induced fatigue causes a decrease of 1.28 and 1.48 times greater in MW than in PM, for the extensors and the flexors, respectively. These observations confirm that it is imperative to assess isokinetic fatigue using the developed MW and not just the PM, according to the different, and therefore non-redundant, information that the MW provides in the context of fatigability resistance evaluation.
Even though the performance of the extensors and the flexors are strongly correlated (PCC around 0.80), the extensors are likely to tire out faster and to a greater extent than the flexors. Two lines of explanation of this phenomenon may be suggested: (1) the greatest absolute force of the extensors compared to the flexors can lead to greater fatigability. Indeed, at equal relative strength, a greater absolute moment seems to accelerate the induction of neuromuscular fatigue [66, 67, 68]. Considering the extensors best performance is 1.78 and 1.62 times greater than the flexors, this factor can, at least partially, explain the observed difference between the extensors and the flexors; (2) as previously explained, the knee flexors produce seldom repetitive maximal concentric contractions which can induce a lower relative strength production due to a lack of specificity of the task. Obviously, a lower relative strength production leads to a lower decrease in performance due to less induced fatigue [69].
The main finding emerging from our study is that whether we consider relative or absolute reproducibility the knee fatigability derived parameters should not be used due to weak reproducibility. Consequently, the diagnostic, conclusions or discussion of a study drawn from the use of the fatigability derived parameters to assessing knee isokinetic fatigability performance appears to be clearly questionable. Indeed, no fatigability derived parameter whatever the quotient of the last
The sample size was a limitation of this study: despite the fact that it is in agreement with the recommendation that reliability protocol should be realized on 15–20 subjects [48, 70], further studies are warranted to confirm the present results over a larger sample size. Another limitation was that this study focused only on the knee joint. As the fatigability resistance evaluations are not only relating to the knee, it would obviously be highly interesting to assess the reliability of the measured and derived parameters in of other joints. Finally, the findings of the present study, as a precautionary measure, should not be generalized to other populations than healthy male subjects.
Conclusions
Our results indicate that all listed fatigability measured parameters (see Table 1 for the complete list) can be used for knee extensors, even if careful consideration must be given to the partial sums at the end of the testing session (#21–#30, #21–#25 & #26–#30). Notably, no fatigability measured parameters can be used for knee flexors, even if the relative reproducibility of the measured parameters for flexor is high. However, the main finding is that no derived parameter based on PM or MW may be used for studying knee extensors or flexors fatigability.
Footnotes
Acknowledgments
The authors wish to thank the Wallonia-Brussels Federation for their assistance in this study.
Conflict of interest
The authors declare that there are no conflicts of interest in undertaking this study.
