Abstract
Quantitative analysis of peptides using internal standards was carried out with the use of matrix-assisted laser desorption/ionization (MALDI) on a time-of-flight mass spectrometer in a ferulic acid matrix. Spectra were collected and averaged over several areas of the sample surface at constant laser energy, and a defocused beam was used to improve reproducibility. A calibration curve was plotted, showing a correlation coefficient of 0.992 and a relative standard deviation of 1.83%. The accuracy and precision of this technique were sensitive to the crystal structure of the matrix and saturation effects of the instrument's detector.
Get full access to this article
View all access options for this article.
