Noise-amplitude spectra were measured for inductively coupled plasma (ICP) atomic emission spectrometry. These spectra indicate that a significant reduction of noise in the emission signal can be obtained by positioning a chimney over the ICP discharge. In this fashion, noise was reduced in the 200–500 Hz region of the frequency spectrum. The chimney had little effect on flicker (1/f) noise.
BelchamberR. M. and HorlickG, Spectrochim. Acta37B, 17 (1982).
2.
DaviesJ. and SnookR. D, J. Anal. At. Spectrom.2, 27 (1987).
3.
HobbsP.SpillaneD. E. M.SnookR. D. and ThorneA. P, J. Anal. At. Spectrom.3, 543 (1988).
4.
WingeR. K.FasselV. A.DeKalbE. L. and EckelsD. E, J. Anal. At. Spectrom.3, 849 (1988).
5.
FurutaN., Presented at the Winter Conference on Plasma Spectrochemistry, San Diego, California (1988), Paper #Th1.
6.
AlkemadeC. Th. J., A Contribution to the Development and Understanding of Flame Photometry, Ph.D. Thesis, University of Ultrecht (1954).
7.
HieftjeG. M. and BystroffR. I, Spectrochim. Acta30B, 187 (1975).
8.
BroekaertJ. A. C.BrushwylerK. R.MonnigC. A. and HieftjeG. M, Spectrochim. Acta45B, 769 (1990).
9.
PressW. H.FlanneryG. P.TeukolskyS. A. and VetlerlingW. T, Numerical Recipes: The Art of Scientific Computing (Cambridge University Press, Cambridge, Great Britain, 1986).
10.
BrushwylerK. R. and HieftjeG. M, paper to appear in Appl. Spectrosc. (1991).