Abstract
Objectives:
Venous filling time (VFT90) is the time taken to reach 90% of the venous volume in the calf. It is recorded by air-plethysmography (APG®) and is assumed to measure global venous reflux duration. However, this has never been confirmed by duplex. The aim of the study was to compare VFT on APG to venous reflux time/duration (RT) measured simultaneously with duplex on the same patients.
Method
Twenty-six consecutive patients, M:
Results
Twenty-five patients are presented. One patient with very mild reflux (VFT90 = 55.9 seconds) had an indeterminate endpoint on duplex and was excluded. The median (range) VFI and GSV diameter was 4.9(1.3–15.5) mL/second and 7(4–17) mm, respectively. The VFT90 and VFT100 both correlated with RT on duplex (Spearman,
Conclusions
This is the first study to compare APG parameters with duplex by performing simultaneous measurements. There was an excellent correlation between the VFT90 versus duplex RT, thereby comparing reverse flow in a single superficial vein against the legs overall venous haemodynamic status. These tests can both be used in the quantification of reflux.
Keywords