Abstract
Background
Dual-energy computed tomography (DECT) has been shown to be useful for subtracting bone or calcium in CT angiography and gives an opportunity to produce a virtual non-contrast-enhanced (VNC) image from a series where contrast agents have been given intravenously. High noise levels and low resolution have previously limited the diagnostic value of the VNC images created with the first generation of DECT. With the recent introduction of a second generation of DECT, there is a possibility of obtaining VNC images with better image quality at hopefully lower radiation dose compared to the previous generation.
Purpose
To compare the image quality of the single-energy series to a VNC series obtained with a two generations of DECT scanners. CT of the urinary tract was used as a model.
Material and Methods
Thirty patients referred for evaluation of hematuria were examined with an older system (Somatom Definition) and another 30 patients with a new generation (Somatom Definition Flash). One single-energy series was obtained before and one dual-energy series after administration of intravenous contrast media. We created a VNC series from the contrast-enhanced images. Images were assessed concerning image quality with a visual grading scale evaluation of the VNC series with the single-energy series as gold standard.
Results
The image quality of the VNC images was rated inferior to the single-energy variant for both scanners, OR 11.5–67.3 for the Definition and OR 2.1–2.8 for the Definition Flash. Visual noise and overall quality were regarded as better with Flash than Definition.
Conclusion
Image quality of VNC images obtained with the new generation of DECT is still slightly inferior compared to native images. However, the difference is smaller with the new compared to the older system.