Abstract
Background
With increasing awareness for radiation exposure, the study of diagnostic accuracy of coronary CT angiography (CCTA) with low radiation dose techniques is mandatory to both radiologist and clinician.
Purpose
To compare diagnostic performance and effective radiation dose between step-and-shoot prospectively ECG-gated and retrospectively ECG-gated with tube current modulation (TCM) CCTA using 128-slice multidetector computed tomography (MDCT).
Material and Methods
We retrospectively evaluated 60 patients who underwent CCTA with either of two different low-dose techniques using 128-slice MDCT (23 patients for step-and shoot-prospectively ECG-gated and 37 patients for retrospectively ECG-gated with TCM CCTA) followed by conventional coronary angiography. All coronary arteries and all segments thereof, except anatomical variants or small size (< 1.5 mm) ones, were included in analysis.
Results
In per-segment analysis, sensitivity, specificity, positive predictive value, and negative predictive value were 91/96%, 95/94%, 75/73%, and 98/99% for step-and-shoot prospectively ECG-gated and retrospectively ECG gated with TCM CCTA, respectively, relative to conventional coronary angiography. Effective radiation dose were 1.75 ± 0.83 mSv, 4.91 ± 1.71 mSv in the step-and-shoot prospectively ECG-gated and retrospectively ECG-gated with TCM CCTA groups, respectively.
Conclusion
The two low-radiation dose CCTA techniques using 128-slice MDCT yields comparable diagnostic performance for coronary artery disease in symptomatic patients with low heart rates.