We report a case of impingement in a metal-on-metal total hip replacement causing both notching of the femoral stem neck and aseptic loosening of the acetabular component. The acetabular component was inserted in excessive anteversion. The femoral stem and acetabular components were retrieved. A larger femoral head was used, and an acetabular component was placed in a less anteverted position. Intra-operative testing through different ranges of movement is recommended to identify potential impingement.
RobbinsGMMasriBAGarbuzDSDuncanCP. Evaluation of pain in patients with apparently solidly fixed total hip arthroplasty components. J Am Acad Orthop Surg2002;10:86–94.
2.
SchmalzriedTPCallaghanJJ. Wear in total hip and knee replacements. J Bone Joint Surg Am1999;81:115–36.
3.
MalikAMaheshwariADorrLD. Impingement with total hip replacement. J Bone Joint Surg Am2007;89:1832–42.
4.
TaninoHItoHHarmanMKMatsunoTHodgeWABanksSA. An in vivo model for intraoperative assessment of impingement and dislocation in total hip arthroplasty. J Arthroplasty2008;23:714–20.
5.
KoyanagiJSakaiTYamazakiTWatanabeTAkiyamaKSuganoN. In vivo kinematic analysis of squatting after total hip arthroplasty. Clin Biomech (Bristol, Avon)2011;26:477–83.
6.
BerendKRLombardiAVJrMalloryTHAdamsJBRussellJHGrosethKL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty2005;20(7Suppl 3):93–102.
7.
KluessDMartinHMittelmeierWSchmitzKPBaderR. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys2007;29:465–71.
8.
BarrackRL. Dislocation after total hip arthroplasty: Implant design and orientation. J Am Acad Orthop Surg2003;11:89–99.
9.
AmstutzHCGrigorisP. Metal on metal bearings in hip arthroplasty. Clin Orthop Relat Res1996;392Suppl:S11–34.
10.
WillertHGBuchhornGHGobelDKosterGSchaffnerSSchenkR. Wear behaviour and histopathology of classic cemented metal on metal hip endoprostheses. Clin Orthop Relat Res1996;329Suppl:S160–86.
11.
McKellopHParkSHChiesaRDoornPLuBNormandP. In vivo wear of three types of metal on metal hip prostheses during two decades of use. Clin Orthop Relat Res1996;329Suppl:S128–40.
12.
DorrLDHiltonKRWanZMarkovichGDBloebaumR. Modern metal on metal articulation for total hip replacements. Clin Orthop Relat Res1996;333:108–17.
13.
IidaHKanedaETakadaHUchidaKKawanabeKNakamuraT. Metallosis due to impingement between the socket and the femoral neck in a metal-on-metal bearing total hip prosthesis. A case report. J Bone Joint Surg Am1999;81:400–3.
14.
McMurtrieAGuhaARWoottonJR. Loose Metasul liner causing partial amputation of the neck of the femoral component. J Arthroplasty2009;24:159.
15.
OndaKNagoyaSKayaMYamashitaT. Cup-neck impingement due to the malposition of the implant as a possible mechanism for metallosis in metal-on-metal total hip arthroplasty. Orthopedics2008;31:396.
16.
LewinnekGELewisJLTarrRCompereCLZimmermanJR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am1978;60:217–20.
17.
TamKHTsangWLChiuSYIpFK. Intermediate-term evaluation of the use of the Wagner acetabular cup for primary total hip replacement in young Chinese patients. Hong Kong J Orthop Surg2004;8:92–6.
18.
MaezawaKNozawaMMatsudaKYuasaTShitotoKKurosawaH. Early failure of modern metal-on-metal total hip arthroplasty using a Wagner standard cup. J Arthroplasty2006;21:522–6.
19.
MurrayDW. Impingement and loosening of the long posterior wall acetabular implant. J Bone Joint Surg Br1992;74:377–9.