Abstract
Introduction
Functional electrical stimulation uses clinician-set parameters to modify stimulation. This study aimed to investigate whether timing parameters in the ODFS Pace functional electrical stimulation device have an effect on the gait of the general population of functional electrical stimulation users who have a foot drop.
Methods
Twelve functional electrical stimulation users with foot drop resulting from upper motor neurone disorders were recruited from the functional electrical stimulation Service in Leeds, UK. A crossover trial design was used, comparing adjusted values of rising ramp, delay and extension. Instrumented gait analysis was carried out to measure ankle dorsiflexion during the swing phase of gait, foot clearance from the ground, and speed of ankle plantarflexion at initial contact. The effect of timing parameters on gait kinematics was studied.
Results
No statistically significant effects on the measured parts of gait were found for any of the timing parameters. Trends were identified in average mid-swing ground clearance and dorsiflexion associated with the delay and rising ramp timing parameters.
Conclusions
Further work in this area should use larger numbers of participants. Based on these results, the effects of ramping and delay would be of particular interest for further study.
Introduction
Drop foot is the inability to raise the foot sufficiently during the swing phase of walking, caused by inadequate control of the dorsiflexor muscles. 1 It is a common functional impairment arising from upper motor neurone lesions, such as stroke, multiple sclerosis and spinal cord injury. 2 On average in human gait, the minimum clearance of the foot above the floor is only 8.7 mm. 3 Drop foot lowers the ground clearance of the foot in swing and so can cause trips and falls even on level ground. Secondary effects on walking include reduced speed, smaller step lengths and lack of confidence.4–6
Functional electrical stimulation (FES) is a functional orthotic intervention, which can be used as an alternative to ankle-foot orthoses to correct drop foot. First proposed by Liberson et al., 7 FES uses an alternating electrical current to stimulate action potentials in neurons. By placing electrodes on the surface of the skin close to the relevant nerves, a changing electrical current between cathode and anode can depolarise nerve fibres, stimulating an action potential which propagates to the relevant muscles. 8
This study used ODFS Pace devices (Odstock Medical Ltd, UK), although some other devices work on similar principles. Activation of the electrical stimulation is electronically controlled, with input from a force-sensitive switch (“footswitch”) beneath the foot, usually the heel. The form of stimulation most commonly used in clinical practice has not significantly changed since the inception of FES,
7
with a train of 3.5–360
Without any timing parameters, stimulation would be active at a single level throughout the swing phase of gait, from heel-rise to footstrike.
9
Timing parameters, normally set by a clinician, alter this stimulation pattern. A graphical representation of the timing parameters in the swing phase of gait is included (Figure 1). The naming convention for timing parameters used in the ODFS Pace is as follows.
Delay delays stimulation initiation by a set number of milliseconds. It can be used to prevent dorsiflexor activation from antagonising plantarflexors during push-off; excessive delay reduces the fraction of swing phase for which stimulation is active. Odstock Medical Ltd recommends a delay of zero in most circumstances. Rising ramp linearly increases the pulse width from zero to the set limit for that device user, within a set number of milliseconds. Ramping begins at the start of stimulation, after heel-rise and any delay. It is thought to increase comfort and prevent stretch reflex activation; excessive rising ramp length may reduce the stimulation amplitude in early swing. Extension extends stimulation beyond footstrike by a set number of milliseconds. It is thought to improve tibial progression and help control the foot to prevent fast forefoot strike during early stance phase; excessive extension may induce fatigue or reduce stability by keeping the foot dorsiflexed during stance. Falling ramp linearly decreases the amplitude of the electrical current within a set number of milliseconds. Ramping begins at the end of stimulation, after footstrike and any extension. It was expected to have a similar but lesser effect to extension, and because of this falling ramp was not studied in this trial. FES stimulation timing parameters.

There have been some previous attempts to change stimulation envelope parameters. Lyons et al. investigated changes in stimulation intensity10–12 in order to better approximate a ‘natural’ stimulation of tibialis anterior. Other authors have studied other aspects of the stimulation envelope, including impulse type 13 and intensity.11,12
A review of the relevant literature has not found any previous work to investigate the effect on gait of timing parameters of the stimulation envelope. This is of clinical interest because optimisation of timing parameters is currently performed by using trial-and-error and individual experience. Research data on the effects of timing parameters’ changes could improve the efficiency of current clinical practice.
This study aimed to investigate whether the delay, rising ramp and extension timing parameters have an effect on the gait of the general population of FES users who have a foot drop. The null hypothesis is that there is no detectable difference in the average gait of users.
Methods
Ethical approval for this study was granted by the West Midlands Research Ethics Committee (approval 16/WM/0132). The study was registered on the clinicaltrials.gov database.
Twelve participants were recruited from the users of the Leeds FES Service. Potential participants were initially identified by clinical staff at the Service, according to the inclusion and exclusion criteria listed here. Written consent was gained at the time of data collection.
Inclusion criteria
Upper motor neurone lesion causing foot drop Currently using ODFS functional electrical stimulator Age 18 or older Attending Leeds FES follow-up clinics
Exclusion criteria
FES user for less than three months Lower limb prosthesis Cannot independently walk for 5 m with walking aids Cannot walk 20 5 -m walks within a 3-h period Uses FES less than once a week
Figure 2 shows the order of events for a participant in the study, together with the drop-out numbers at each stage.
Flowchart showing the order of events for a participant in the study.
Population characteristics.
Recruited participants were asked to visit the Leeds gait analysis laboratory at Chapel Allerton Hospital in Leeds, UK. The Leeds gait analysis laboratory uses 8 Vicon MX-T40 cameras, 14 with Vicon Nexus 1.8.5 software. Retroreflective markers (14 mm diameter) are placed using the Helen Hayes markerset.
FES settings for each set of walks, where ‘user’ means the user’s usual setting.
The order in which settings are tested was non-sequential; this is to reduce the confounding factors of any short-term carry-over or training effect. Single-blinding was used, as double-blinding was not practicable due to limitations in staff time.
Data processing was completed initially in Vicon Polygon 4.2 software 14 by an experienced gait analyst. One stride was chosen as representative from each set of trials, and events in this stride were marked for both left and right: initial foot-strike, foot-off, and final foot-strike.
Data analysis was performed using Python 3.5.5. 15 Marker trajectories and joint angles for each marked stride were exported to comma-separated-value format, and data visualisation was performed using the matplotlib package for Python.
Results were obtained for the ankle dorsiflexion, ground clearance, walking speed and speed of plantarflexion at initial contact. The results of interest were further analysed; these are ankle dorsiflexion and ground clearance for the delay and ramp conditions, and the plantarflexion speed at initial contact. Ground clearance of the toe at mid-swing was calculated as the minimum height of the toe in the central 60% of the swing phase. Plantarflexion speed was calculated as the fastest vertical speed of the toe in the first 20% of the stance phase.
Histograms showed approximately normal distribution of data. Tests for statistical significance were conducted on results for mid-swing ground clearance and foot plantarflexion using one-way analysis of variance (ANOVA). Effect size was calculated as
Results
Figures 3 and 4 show the mid-swing clearance of the foot from the ground for delay and rising ramp conditions. Figure 5 shows the speed of ankle plantarflexion after initial contact, for the extension conditions.
Box plot showing all participants’ mid-swing clearance from ground at different delays. Box plot showing all participants’ mid-swing clearance from ground at different rising ramps. Box plot showing speed of all participants’ ankle plantarflexion after initial contact, at different extensions.


There was a marked divide in participants between those with a self-selected walking speed over 0.5 m/s (min 0.65, max 0.8 m/s average speed) and those with a walking speed below 0.5 m/s (min 0.2 max 0.4 m/s average speed). Sub-group analysis was performed on faster (n = 6) and slower (n = 6) participants. Figures 6 and 7 show the differing effects of delay on mid-swing foot clearance in slower and faster walkers, respectively. There was no difference between fast and slow walkers in average plantarflexion speed.
Box plot showing slow participants’ mid-swing clearance from ground at different delays. Box plot showing fast participants’ mid-swing clearance from ground at different delays.

Basic statistics for each timing parameter.
Discussion
All participants were able to complete an effective gait with all timing parameters used. Calculations of effect sizes show small to negligible effects on walking speed, clearance and plantarflexion speed from most parameter changes. Two of the delay parameters gave had a moderately sized effect on plantarflexion speed, using Cohen’s suggested medium effect size of 0.5–0.8. 16 No results were statistically significant and the limitations that may have caused this are discussed below.
A low number of participants who were successfully recruited, with data collected from only 12 participants. The low number is partly due to time constraints and partly due to limited uptake amongst potential participants. The study had broad inclusion criteria intended to identify adults with a foot drop who were regularly using an ODFS who were able to complete the study protocol. However that population with a foot drop is not heterogeneous. The study sample is likely to have contained participants with various degrees of lower-limb muscle spasticity. Ramping of stimulation is thought to help to reduce the stimulation of the reflex seen in spasticity, as it reduces the velocity of dorsiflexion. 9 If this were the case, for the rising ramp parameter one would expect participants with spasticity to have reduced dorsiflexion with smaller rising ramps, as spastic reflex of the plantarflexors would have reduced their ankle range of movement.
Fatigue could have an effect on results, as the order of walks was the same for each participant (i.e. not randomised). The first three participants to take part were timed walking 10 m with their usual FES settings, at the beginning and end of the data collection session. These participants slowed by 6%,
The limited resolution of timing parameters (50 ms steps) precluded normalisation of parameters to our participants’ walking speed or stride time. The effect of timing is dependent on the speed of walking because a given time period takes a different fraction of the gait cycle depending on the length of that cycle. For example, the longest rising ramp time in the study was 300 ms: this corresponds to about half of the swing phase in the slowest participants but nearly all of the swing phase in the fastest. Sub-group analysis of faster and slower walkers allowed us to see whether this difference in relative timings had any effect on the data recorded.
As described in the methods section, falling ramp was not studied. Future studies might find this parameter to be of interest.
Conclusions
This study has assessed the effect of FES timing parameters on the gait of FES users. The results were not statistically significant and so the null hypothesis has not been disproven; however, it is hoped that these results may be instructive for further research. Any further work in this area might be useful with larger numbers of participants. More restrictive inclusion criteria, perhaps including walking speed, may help in future studies.
Calculated effect sizes show that most timing parameter changes have small or negligible effects on the walking parameters measured here. The moderate effect sizes seen in ankle plantarflexion speed at initial contact, with changes in delay, suggest that this may be a potential avenue for future studies. Further analysis of these specific effects with a larger cohort of participants would be needed before recommendations could be given to clinicians.
Footnotes
Declaration of conflicting interests
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding
The author(s) received no financial support for the research, authorship, and/or publication of this article.
Guarantor
KT.
Contributorship
SM and KT conceived the study, gained ethical approval. SM researched literature, developed the protocol, analysed the data, and wrote the first draft of this manuscript. Both SMs and LM were involved in recruitment, data collection and analysis. All authors reviewed and edited the manuscript and approved the final version of the manuscript.
Acknowledgements
This work was completed as part of a MSc research project for King’s College London. The findings within this paper were previously presented at the 21st Annual Conference of the International Functional Electrical Stimulation Society, IFESS, London, 2017. We would like to thank Adam Shortland for his guidance on reporting these results.