Abstract
Background:
Artificial urinary sphincter (AUS) activation is usually deferred until 6 weeks following insertion. However, this timeframe for activation is based on traditional dogma. This study aims to assess the feasibility, efficacy and safety of earlier activation of AUS.
Methods:
A prospective database was established on consecutive patients undergoing AUS implantation (AMS-800 Boston Scientific) from March 2019 to March 2022 by a single fellowship-trained urologist. Early activation was defined as <30 days from the date of implantation.
Results:
Twenty-one patients were included in the study with a median follow-up of 34 (range: 3–96) months. The most common indication for AUS insertion was post-prostatectomy stress urinary incontinence (n = 19, 90%). All 21 patients had their AUS successfully activated and used before 30 days, with a median of 14 (range: 9–28) days. Median (range) pre-operative continence pad numbers were 3 (2–6) and 0 (0–1) at 4 weeks post-operatively. Median (range) pad weight per day was 600 g (190–1310 g) pre-operatively and 6 g (0–6 g) at 4 weeks post-operatively. There were no adverse events associated with early activation.
Conclusion:
To our knowledge, this is the first description of the early activation of the AMS-800 AUS device. Our preliminary experience indicates early activation is safe and acceptable for patients.
Level of evidence:
II
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.