Abstract
Background:
This study examines how shield-triggered autoinjectors (AIs), for subcutaneous drug delivery, affect injection depth. It focuses on shield size and applied force, parameters that could potentially lead to inadvertent intramuscular (IM) injections due to tissue compression.
Method:
A blinded ex-vivo study was performed to assess the impact of shield size and applied force on injection depth. Shields of 15, 20, and 30 mm diameters and forces from 2 to 10 N were investigated. The study involved 55 injections in three Landrace, Yorkshire, and Duroc (LYD) pigs, with injection depths measured with computed tomography (CT). An in-vivo study, involving 20 injections in three LYD pigs, controlled the findings, using fluoroscopy (FS) videos for depth measurement.
Results:
The CT study revealed that smaller shield sizes significantly increased injection depth. With a 15 mm diameter shield, 10 N applied force, and 5 mm needle protrusion, the injection depth exceeded the needle length by over 3 mm. Injection depth increased with higher applied forces until a plateau was reached around 8 N. Both applied force and size were significant factors for injection depth (analysis of variance [ANOVA],
Conclusions:
The study demonstrates that shield size has a greater impact on injection depth than the applied force. While conducted in porcine tissue, the study provides useful insights into the relative effects of shield size and applied force. Further investigations in humans are needed to confirm the predicted injection depths for AIs.
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.