Abstract
Background:
The aim of this study was to evaluate a newly developed system for insulin delivery incorporating a multifunctional blood glucose meter and a remotely controlled insulin pump (ACCU-CHEK® Combo system) in established pump users with type 1 diabetes. The technology was assessed both from device performance and subject usability perspectives.
Method:
A multicenter, prospective, single group study was carried out in five centers in the Netherlands and four centers in the United Kingdom for more than 6 months. The study was divided into two phases: Phase 1 (4 weeks) for device validation purposes and phase 2 (22 weeks) to observe the impact of the system on metabolic control, patient satisfaction [using the Diabetes Treatment Satisfaction Questionnaire (DTSQ)] and device safety.
Results:
Eighty subjects completed the planned study period. There were no unexpected device errors. Treatment satisfaction was high at baseline and further increased to study end (DTSQ change version: Sum score, 10.6 ± 7.2; scale score range, −18 to +18, p < 0.0001). Hemoglobin A1c improved continuously over time, from 7.9% (±0.9%) to 7.7% (±0.8%) at month 3 (p < 0.001) and 7.6% (±0.8%) at month 6 (p < 0.0001). The frequency of severe hypoglycemia was 0.08 per patient years. There was no case of ketoacidosis.
Conclusions:
The new system was evaluated by experienced continuous subcutaneous insulin infusion users as safe in daily practice and associated with favorable treatment satisfaction and a modest improvement in glycemic control.