Abstract
Background
Distal biceps tendon repair using endobutton fixation has shown the best biomechanical results in terms of pullout strength. Here, we compared Sethi’s enhanced tension adjustable endobutton technique known as the “tension slide technique” to a new knotless endobutton fixation technique without a post-fixation screw. Our new approach is as effective as the tension slide technique in terms of pullout strength and gapping after early mobilization.
Methods
A biomechanical cadaveric study with 16 paired arms was performed. With the radius held in place, the distal biceps tendon was loaded at 100 N for 500 cycles and the load was then increased until failure. Gapping after loading cycles and maximum load to failure were recorded and compared.
Results
Median bone-tendon gapping was 5.77 mm (interquartile range (IQR) 4.84–9.11) for tension slide technique and 4.72 mm (IQR 1.77–6.16) for the knotless fixation (p = 0.047). Median load to failure was 257.87 N (IQR 222.07–325.35) in the tension slide technique group and 407.78 N (IQR 358.54–485.20) in the knotless group (p = 0.047).
Discussion
The knotless endobutton provides better pullout strength and elongation results compared to the tension slide technique without the use of an interference screw, allowing early mobilization in order to faster return to daily living activities.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.