Abstract
Background
Buried free flaps represent a reconstructive challenge concerning monitoring of vitality, which is fundamental for an early detection of flap failure and prompt surgical salvage. Many flap monitoring techniques have been developed over time, and there is still no consensus concerning the best monitoring technique for buried reconstructions.
Methods
We performed a systematic review of the literature on NIRS monitoring for head and neck free flaps. Moreover, we presented a case of orbital reconstruction through a buried free myo-fascial anterolateral tight flap (ALTF) in which postoperative monitoring was performed by means of NIRS.
Results
Four studies were included with a total of 200 monitored head and neck free flap reconstructions. Flap survival was reported in 96.5% of studies (n = 193/200) with a 3.5% of total flap failure rate (n = 7/200). We monitored the buried myo-fascial ALTF for 7 post-operative days measuring a regional oxygen saturation (rSO2) ranging from 55% to 72% (mean = 66%).
Conclusions
This device appeared to be an efficient choice for monitoring buried flaps, thanks to its ability to measure tissue perfusion deep under the skin, to the continuous availability of recorded data on the monitor, and to its low impact on the patient. Further prospective studies are advised in order to standardize this monitoring technique and define warning values.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.