Abstract
We report a case of a 77-year-old male with pancreatic cancer and thrombocytopenia who presented with acute stroke symptoms and underwent successful endovascular thrombectomy. During femoral artery closure, an 8-F Angio-Seal device fractured, leaving catheter fragments in the right femoral artery, confirmed on imaging. Device analysis revealed oxidation-induced brittleness, leading to sheath fragmentation. This case highlights a rare complication of Angio-Seal devices, suggesting potential material vulnerabilities that warrant further investigation.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.