Abstract

Radiotherapy planning, delivery and junction dose verification remain exigent for Cranio
Spinal Irradiation (CSI) in medulloblastoma patients. This study aims to evaluate high
precision techniques such as Intensity Modulated Radiation Therapy (IMRT), Rapid Arc Therapy
(RA) with and without flattening filter (FF) on the basis of dosimetric analysis. Five
patients treated with jagged junction Intensity Modulated RadioTherapy (IMRT) using dynamic
Multi Leaf Collimators (MLC) were randomly selected for this retrospective study. IMRT, Rapid
Arc (RA) plans were simulated in the same CT data set with and without flattening filter.
Total dose prescribed was 28.80 Gy in 16 fractions. An evaluation criterion of 98% of PTV
receiving 100% of the prescription dose was followed in all plans. Twenty treatment plans with
260 Dose Volume Histograms (DVHs) was created. Dosimetric parameters such as Dmax,
Dmin, Dmean, V95%, V107%, CI for PTV and
Dmax, Dmean, V80%, V50%, V30%,
V10% for Organs At Risk (OAR) were extracted from DVHs. Treatment delivery
efficiency was also evaluated for total Beam On Time (BOT). FFF Rapid Arc therapy