Abstract
Purpose:
To define analysis methods using radiofrequency (RF) data from intravascular ultrasound (IVUS) to characterize thrombus so that the success or failure of thrombolysis can be predicted.
Methods:
Experimentation was done in 2 phases: first, 40 clots created from platelet-rich (n = 20) and platelet-poor (n = 20) plasma were imaged with 20 and 30-MHz IVUS probes. The digitized RF signals were analyzed to determine the attenuation and slope of attenuation characteristics as a reference standard for the second part of the study. In this phase, 20 perfusion trials were performed with 10 platelet-rich and 10 platelet-poor samples in an in vitro arterial perfusion model; the composition of the samples was blinded during interrogation with IVUS. Alteplase (0.1 mg) was then infused directly into the thrombus for 1 hour; restoration of flow and amount of lysis were quantified.
Results:
Significant differences were detected for the integrated attenuation value (p<0.001) between platelet-poor and platelet-rich clots interrogated with a 30-MHz probe, but no differences were noted using the 20-MHz catheter. However, both transducers were capable of differentiating platelet-rich and platelet-poor clots using a slope of attenuation analysis, which was successful in identifying all 20 (100%) of the blinded clots in the in vitro model. The speed and completeness of lysis were significantly higher (p<0.05) for platelet-rich (78%) than platelet-poor (21%) clots.
Conclusions:
IVUS preprocessed imaging is capable of differentiating various types of thrombus. This information can be used to predict the success or failure of pharmacological lysis techniques.
Keywords
Get full access to this article
View all access options for this article.