Abstract
Aim:
To investigate the effect of femoral head size on blood metal-ion levels caused by taper corrosion in metal-on-polyethylene total hip arthroplasty, comparing 36- to 44-mm heads with 32-mm heads.
Methods:
In a randomised, controlled, single-blinded trial, 96 patients were allocated to receive either a 32-mm metal head or the largest possible metal head (36–44 mm) that could be accommodated in the thinnest available vitamin E, cross-linked polyethylene insert. Blood metal ion levels were collected at 1- and 2-year follow-ups.
Results:
At 1-year, metal-ion levels did not differ between the groups. The median (interquartile range) blood-ion levels for the 32-mm versus the 36- to 44-mm group were 0.11 µg/L (0.08–0.15) versus 0.12 µg/L (0.08–0.22),
Conclusions:
The use of the largest possible metal head (36–44 mm) compared to a 32-mm head in metal–on-polyethylene bearings does not appear to elevate blood metal-ion levels up to 2 years postoperatively. As taper corrosion is probably time-dependent, longer-term reports are needed to evaluate the association between large metal heads and blood metal ion levels.
Trial registration: ClinicalTrials.gov (reg. ID NCT0231 6704)