Abstract
Background
Impella CP® is a percutaneous left ventricular assist device that requires a heparin–dextrose purge solution. The manufacturer recommends heparin 50 units/mL, but supratherapeutic anticoagulation has been observed with this concentration.
Objective
The purpose of this evaluation was to observe the efficacy and safety of a low-dose heparin-based purge solution (25 units/mL in dextrose 20%). The primary outcome evaluated percentage of activated clotting times (ACTs) below therapeutic range. Secondary objectives included evaluating the incidence of device thrombosis and rate of heparin-induced thrombocytopenia (HIT). Platelet trends were characterized.
Methods
A single-site retrospective review was conducted for all adults with the Impella CP from January 2015 to December 2017.
Results
A total of 18 patients were included. The percentage of ACT readings within goal of 160 to 200 seconds was 49%, and 38% of readings were subtherapeutic. Per BARC bleeding criteria, 22% (n = 4) patients experienced class IIIa bleeding and 39% (n = 7) experienced class II bleeding. Though 4 (22%) patients were tested for HIT, no patients were positive. Patients showed universal reductions in platelet counts.
Conclusions
The use of a low-dose anticoagulation protocol of heparin 25 units/mL in dextrose 20% as needed warrants further evaluation.