The feasibility of non invasive, automatic intradialytic measurement of effective clearance is demonstrated by a mathematical derivation. In a practical setting the conductivity difference between dialysate inlet and outlet of the dialyzer is measured for two different dialysate inlet concentrations. This allows for the calculation of the electrolyte dialysance, which is equal to the effective urea clearance, provided dialysate flow, blood flow and blood electrolyte composition remain unchanged during the measuring interval
GotchF.A., SargentJ.A.A mechanistic analysis of the National Cooperative Dialysis Study (NCDS).Kidney Int1985; 28: 526–34.
2.
PolascheggH-D, HusarD.Dialysevorrichtung mit geregelter Zusammensetzung der Dialysierloesung. European Patent 0097366, priority date 21.06.1982.
3.
SargentJ.A., GotchF.A.Principles and biophysics of dialysis. In: DrukkerW., ParsonsF.M., MaherJ.F., eds. Replacement of Renal Function by Dialysis, Second Edition.Boston: Martinus Nijhoff Publishers, 1983; 53–96.
4.
PolascheggH-D. Verfahren zur in vivo Bestimmung von Parametern der Hämodialyse. German Patent Application DE 39 38 662 A1, priority date 21.11.1989.
5.
SteilH., PolascheggH-D. (to be published).
6.
PolascheggH-D. Vorrichtung zur Bestimàmung des behandelten Blutvolumens bei der Hämodialyse. European Patent Application No: 91108005. Priority date; 17.05.1991.
7.
SchneditzD., PolascheggH., LevinN.W.Cardiopulmonary recirculation in dialysis; an under-recognized phenomenon.ASAIO Journal1992; XXXVIII: M194–6.
8.
AlbersJ.R., SmithJ.M.A conductivity technique for rapid measurement of in vitro dialyzer performance.ASAIO Trans1965; XI: 161–4.
9.
BoagJ., VlchekD.Clearance testing of dialyzers using conductivity.Contemporary Dialysis & Transplantation1987; 8: 11.
10.
GotchF.A.Models to predict recirculation and its effect on treatment time in single-needle dialysis. In: RingoirS., VanholderR., IvanovichP., eds. First International Symposium on Single-Needle Dialysis.Cleveland: ISAO Press, 1984; 47–62.