Abstract
A 5-month-old male was treated with left ventricular assist device due to cardiac failure secondary to dilated cardiomyopathy. The patient developed acute severe intravascular hemolysis with methemoglobinemia and renal failure, related to a mechanical problem due to pump cylinder misalignment. Secondary severe methemoglobinemia has not been previously described in patients undergoing ventricular assist device. Early detection of the signs and symptoms of hemolysis is crucial to prevent further complications.
Get full access to this article
View all access options for this article.