Abstract
Background
The routine follow-up of cardioverter defibrillators (CD) is a time-consuming procedure.
Aim of the Study and Methods
The present study was a prospective randomized cross-over evaluation on the clinical usefulness of a specific semi-automatic software algorithm (Quick Check) for CD follow-up, available in CPI Guidant systems (CD and programmer). Time-saving, while ensuring all the required data and patient safety, was evaluated in a large group of patients (105), recruited in different centers. In the same session and under a physician's supervision all patients underwent a follow-up with the aid of Quick Check or a standard follow-up, in a randomized sequence. Each patient served as his own control.
Results
In the overall population of 105 patients, the time spent for follow-up was reduced by Quick Check from 186±105 sec to 106±67 sec (p<0.0001) (43% reduction). The reduction in time spent for follow-up with Quick Check was the same (43% reduction) in patients with detected episodes (n=38) (from 241±144 sec to 138±95 sec (p<0.0001)) and in patients without detected episodes (n=67) (from 154±52 sec to 88±34 sec (p<0.0001)). No adverse events or deletion of potentially useful data was detected by the supervising physician.
Conclusions
Use of a specific software algorithm for routine follow-up of implanted CDs allows a significant shortening of the time spent on routine follow-up, thus reducing costs. The supervision of a physician is a guarantee of patient safety.
Get full access to this article
View all access options for this article.