Abstract
Background:
Osteotomies at an angle of 45° to the long axis of the glenoid were originally used in a cadaveric model to simulate the bone loss that can occur clinically in anterior instability of the shoulder. However, this type of glenoid defect is not consistent with the usual clinical scenario, in which bone loss occurs nearly parallel (at 0°) to the long axis of the glenoid.
Purpose:
Our objectives were to compare the amount of glenoid bone loss measured after a 45° glenoid osteotomy with that after a 0° osteotomy and to determine differences in bone loss measurement from 2 different posterior shoulder portals.
Study Design:
Controlled laboratory study.
Methods:
Glenoids of 14 embalmed cadaveric shoulders (mean age, 81 years; range, 56–90) were mounted in a custom shoulder holder, and 2 posterior portals (2 and 3 o'clock) were fixed into place. The area of a best-fit circle of the inferior portion of the glenoid was digitally calculated, and 2 sequential osteotomies of 12.5% and 25% of anteroinferior glenoid bone loss area were created. Two different types of osteotomies were created: group 1, “inverted-pear” bone loss (45° to the long axis of the glenoid); and group 2, “clinical” bone loss osteotomy (0° to the long axis of the glenoid). Measurements of bone loss were performed based on the bare spot method from 2 simulated posterior portals at 2 and 3 o'clock using a calibrated probe and digital calipers. The osteotomy was measured in 3 different locations (upper, middle, and lower thirds).
Results:
In the 12.5% bone loss model, bone loss measurements for both groups were significantly higher than expected (22.2%–23.1 % in group 1,17.4%–17.9% in group 2;
Conclusion:
Glenoid bone loss determination in a 45° osteotomy model significantly overestimates the amount of true glenoid bone loss. However, in a 0° clinical bone loss simulation model, the arthroscopic bare spot method of bone loss determination was sufficiently accurate at all 3 areas (upper, middle, and lower third) of bone loss. Both the 2-o'clock and 3-o'clock posterior portals were accurate to determine the amount of glenoid bone loss as referenced from the bare spot.
Clinical Relevance:
Arthroscopic determination of glenoid bone loss is more accurate than what has been previously described with the 45° simulation model. Measurement of glenoid bone loss from either the 2-o'clock or 3-o'clock posterior portal is accurate in a clinical bone loss model.