Abstract
Background
Photon-counting detector computed tomography (PCD-CT) is a groundbreaking technology with promising results for visualization of small bone structures.
Purpose
To analyze the delineation of the thoracic spine in multiplanar reconstructions (MPR) on PCD-CT compared to energy-integrating detector (EID)-CT.
Material and Methods
Two euthanized mice were examined using different scanners: (i) 20-slice EID-CT and (ii) dual-source PCD-CT at various CTDIVol values. Readers evaluated the thoracic spine and selected series with best visualization among signal-to-noise ratio (SNR)-matched pairs.
Results
SNR was significantly higher in PCD-CT reconstructions (Br68) and lower in Hr98 reconstructions compared to EID-CT. Bone detail visualization was superior in PCD-CT (especially in Hr98 reconstructions) compared to EID-CT.
Conclusion
MPR on a PCD-CT had a higher SNR and better bone detail visualization even at lower radiation doses compared to EID-CT. PCD-CT with bone reconstructions showed the best delineation of small bone structures and might be considered in clinical routine.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.