Abstract
Background
Dynamic myocardial computed tomography perfusion (CTP) is a novel imaging technique that increases the applicability of CT for cardiac imaging; however, the scanning requires a substantial radiation dose.
Purpose
To investigate the feasibility of dose reduction in dynamic CTP by comparing all-heartbeat acquisitions to periodic skipping of heartbeats.
Material and Methods
We retrieved imaging data of 38 dynamic CTP patients and created new datasets with every fourth, third or second beat (Skip1:4, Skip1:3, Skip1:2, respectively) removed. Seven observers evaluated the resulting images and perfusion maps for perfusion deficits. The mean blood flow (MBF) in each of the 16 myocardial segments was compared per skipped-beat level, normalized by the respective MBF for the full dose, and averaged across patients. The number of segments/cases whose MBF was <1.0 mL/g/min were counted.
Results
Out of 608 segments in 38 cases, the total additional number of false-negative (FN) segments over those present in the full-dose acquisitions and the number of additional false-positive cases were shown as acquisition (segment [%], case): Skip1:4: 7 (1.2%, 1); Skip1:3: 12 (2%, 3), and Skip1:2: 5 (0.8%, 2). The variability in quantitative MBF analysis in the repeated analysis for the reference condition resulted in 8 (1.3%) additional FN segments. The normalized results show a comparable MBF across all segments and patients, with relative mean MBFs as 1.02 ± 0.16, 1.03 ± 0.25, and 1.06 ± 0.30 for the Skip1:4, Skip1:3, and Skip1:2 protocols, respectively.
Conclusion
Skipping every second beat acquisition during dynamic myocardial CTP appears feasible and may result in a radiation dose reduction of 50%. Diagnostic performance does not decrease after removing 50% of time points in dynamic sequence.
Keywords