Abstract
Background
The cerebral aqueduct is a central conduit for cerebrospinal fluid (CSF), and non-invasive quantification of CSF flow in the aqueduct may be an important tool for diagnosis and follow-up of treatment. Magnetic resonance (MR) methods at clinical field strengths are limited by low spatial resolution.
Purpose
To investigate the feasibility of high-resolution through-plane MR flow measurements (2D-PC) in the cerebral aqueduct at high field strength (7T).
Material and Methods
2D-PC measurements in the aqueduct were performed in nine healthy individuals at 7T. Measurement accuracy was determined using a phantom. Aqueduct area, mean velocity, maximum velocity, minimum velocity, net flow, and mean flow were determined using in-plane resolutions 0.8 × 0.8, 0.5 × 0.5, 0.3 × 0.3, and 0.2 × 0.2 mm2. Feasibility criteria were defined based on scan time and spatial and temporal resolution.
Results
Phantom validation of 2D-PC MR showed good accuracy. In vivo, stroke volume was −8.2 ± 4.4, −4.7 ± 2.8, −6.0 ± 3.8, and −3.7 ± 2.1 µL for 0.8 × 0.8, 0.5 × 0.5, 0.3 × 0.3, and 0.2 × 0.2 mm2, respectively. The scan with 0.3 × 0.3 mm2 resolution fulfilled the feasibility criteria for a wide range of heart rates and aqueduct diameters.
Conclusion
7T MR enables non-invasive quantification of CSF flow and velocity in the cerebral aqueduct with high spatial resolution.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.