The performance of the SMA 12/micro autoanalyser in clinical use in a children's hospital is described. Details of the precision and accuracy over a period of 134 operating days are given, and some of the factors which influence the results are discussed. The slow operating speed and analytical difficulties on the sodium, potassium, and cholesterol channels made the analyser unsuitable for routine use in this laboratory.
References
1.
BroughtonP. M. G.AnnanW.Measurement of analytical precision in clinical chemistry. Clin. chim. Acta, 32 (1971) 433.
2.
BroughtonP. M. G.ButtolphM. A.GowenlockA. H.NeillD. W.SkentelberyR. G.Recommended scheme for the evaluation of instruments for automatic analysis in the clinical biochemistry laboratory. J. clin. Path.22 (1969) 278.
3.
BauerS.DeVinoT.MatthewsS.SMA 12/micro: An evaluation based on parallel analysis of patients specimens with an SMA 12/60. Advances in automated analysis. Technicon International Congress, 1970, Vol 1, 41.
4.
CrerandR. F.HochstrasserH.MorgansternS.YoonY. R.VaradyR.The SMA 12/micro. Advances in automated analysis. Technicon International Congress, 1970, Vol 1, 37.
5.
HabigR. L.An evaluation of eleven months’ operating experience with the SMA 12/micro autoanalyser. Automation in Analytical Chemistry. Technicon International Symposium, (1971) p. 3.
6.
HenryR. J.ChiamoriN.GolubO. J.BerkmanS.Revised spectrophotometric methods for determination of glutamic oxaloacetic transaminase, glutamicpyruvic transaminase, and lactic acid dehydrogenase. Amer. J. clin. Path.34 (1960) 381.
7.
KingJ.Practical Clinical Enzymology, Van Nostrand, London, 1965, p. 202.
8.
KingE. J.WoottonI. D. P.Microanalysis in Medical Biochemistry, Churchill, London, 1956, p. 81.
9.
LoganR. W.TweedieA. K.An Evaluation of the SMA 12/micro autoanalyser in the clinical biochemistry laboratory. Clin. chim. Acta, 44 (1973) 73.
10.
MeitesS.Reproducibly simulating haemolysis, for evaluating its interference with chemical methods. Clin. Chem.19 (1973) 1319.
11.
PriceC. P.WoodmanD. D.An improved autoanalyser technique for the determination of serum alkaline phosphatase. Clin. chim. Acta35 (1971) 265.