Abstract
Silicon wafer-based internal reflection elements (IREs) present many practical advantages over the prisms conventionally used for attenuated total reflection (ATR) spectroscopy in the infrared. We examine two methods of using minimally prepared IREs that have appeared in the literature, edge-coupled (EC) and prism-coupled (PC), in conjunction with a liquid flow cell. Polarization measurements show that radiation entering the PC-IRE becomes depolarized due to stress-induced birefringence, and transmission through the edge of the EC-IRE also affects the polarization state. Quantification of the noise and a calibration using a series of sodium acetate solutions show the sensitivity of the PC-IRE outweighs the lower noise obtainable with the EC-IRE.