Standard insulin infusion sets (IISs) are to be replaced every 2 to 3 days to avoid complications and diabetic ketosis due to set failure. This pivotal trial evaluated the safety and performance of a new extended-wear infusion set (EIS) when used for 7 days by adults with type 1 diabetes (T1D).
This single-arm, nonrandomized trial enrolled adults (18–80 years of age) with T1D, who used their own MiniMed™ 670G system with insulin lispro or insulin aspart and the EIS for up to 7 days, across 12 consecutive wears. Safety endpoints included incidence of serious adverse events (SAEs), serious adverse device effects (SADEs), unanticipated adverse device effects (UADEs), severe hypoglycemia (SevHypo), severe hyperglycemia (SevHyper), diabetic ketoacidosis (DKA), and skin infection. The EIS failure rate due to unexplained hyperglycemia (i.e., suspected occlusion), the overall EIS survival rate, glycemic control outcomes (i.e., A1C, mean sensor glucose and time spent in established glucose ranges), total daily insulin delivered, and satisfaction with the EIS were determined.
The intention to treat population (
This investigation demonstrates that the EIS, when worn for up to 7 days, was safe and rated with high satisfaction, without adversely affecting glycemic control in adults with T1D.
Clinical Trial Registration number: NCT04113694 (