Abstract
The purpose of this study is to compare the dose distribution, monitor units (MUs) and radiation delivery time between volumetric-modulated arc (VMAT) and fix-beam intensity modulated radiotherapy (FB-IMRT) in skull-base and non-skull-base head and neck cancer (HNC). CT datasets of 8 skull-base and 7 non-skull-base HNC were identified. IMRT and VMAT plans were generated. The prescription dose ranged 45–70 Gy (1.8–2.2 Gy/fraction). The VMAT delivery time was measured when these plans were delivered to the patients. The FB-IMRT delivery time was generated on a phantom. Comparison of dose-volume histogram data, MUs, and delivery times was performed using T-test. Our results show that both plans yield similar target volume coverage, homogeneity, and conformity. In skull-base cases, compared to FB-IMRT, VMAT generated significantly smaller hot-spot inside PTV (2.0% vs. 4.5%,