Abstract
This paper proposes a novel artificial anal sphincter system (AASS) for severe fecal incontinence. The AASS is composed of an artificial anal sphincter (AAS), an external transcutaneous energy transmission system (TETS), and an external control device. The AAS is composed of a cuff, a micropump, a reservoir, and a remote control device. It is designed to be implanted into the body of the patient. The function of the AAS is to open and close the patient's natural anus. Patients suffering from loss of their natural sphincter lose rectal sensation and are thus unable to perceive imminent fecal incontinence. In order to restore rectal sensation, a pressure sensor in the AAS cuff is designed to detect pressure in the colon. The pressure reflects the present quantity of colon contents, allowing patients to control the AAS to open or close the anus according to the pressure. The TETS is designed to provide electrical energy to the implanted AAS without wire connections. The external control device is designed to receive the pressure information from the AAS and send the patient's command to the implanted device. This paper provides a thorough discussion of the design of the novel AASS and describes the performance of the AASS when tested in vivo on two Beagle dogs who were chosen to be the subjects for receiving the implant. The experimental results verified that the performance of the AASS met the functional requirements it was designed for; however, the trial also revealed some challenges to be further studied.